Hologic ATEC SAPPHIRE - Stereotactic Adapter Cleaning

Hologic ATEC SAPPHIRE
36 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
33
FIGURE 8: ATEC STX-2F FIGURE 9: ATEC STX-FISCHER
Cleaning Instructions
Follow these instructions to clean the Stereotactic Adapter:
1. Upon completion of the biopsy procedure, immediately rinse the adapter with warm water.
2. Spray the adapter with a cleaner.
CAUTION: The following products are NOT RECOMMENDED as they may damage the exposed components:
a. Bleach based cleaning agents.
b. Hydrogen Peroxide
c. Any cleaners/disinfectants with a pH of less than 4 or greater than 10
3. Allow the necessary time for the chosen cleaner to disinfect the components.
4. Rinse the adapter with warm water.
5. Wipe the adapter dry with a cloth or a paper towel.
6. Cycle hub retaining clamps, handpiece retaining clamp, cock and re adapter.
7. Re-Clean the stereotactic adapter if any component does not perform properly.
Note: There are many congurations of stereotactic systems available. The ATEC stereotactic adapter congurations
depicted in this manual may require a unique tting to connect to your particular manufacturer/model. Please contact
Hologic Customer Support at 1-877-887-8767 if you believe this accessory, along with the appropriate instructions, was
omitted. If you are an international customer, please contact your distributor directly with questions, comments, and/or
technical service issues.
Adapter Adapter
Needle Guide Holder Needle Guide Holder