• After connecting to a probe, the UFU must be turned ‘off’
then back ‘on’ before attempting to connect to a new
probe.
Probe
• DO NOT activate the Probe until it is ready to be used. It is
designed to be used immediately after activation.
• Keep a spare Probe at the ready in case of a failure, or
of exceeding its 100-minute life after activation, or of
exceeding its 4-hour life window.
• DO NOT reuse the Probe. It can be used on a single patient
in a single surgical procedure only. Reuse poses a risk of
cross-infection and cross-contamination.
• DO NOT attempt to re-sterilize the Probe.
• DO NOT drop the Probe. It may damage the detector
element.
• DO NOT strike the Probe tip against a hard surface. It may
damage the detector element.
• DO NOT simultaneously use the probe together with
an electro-surgery device. It can disrupt the detector
and produce spurious counts.
• DO NOT touch the probe to an energized electro-
surgery device. It can damage the probe.