10
4.
USING THE SYSTEM
Setting up the UFU (to be done prior to starting
the procedure)
1.
Turn on the UFU by pressing down on the ON-OFF control
switch located on its top-left edge for approximately 5 seconds,
until the boot screen appears.
If only the battery-charge ICON appears, the battery is
fully depleted and the UFU must be recharged prior to use.
NOTE: A full charge for a fully-discharged UFU takes about three
hours.
Unless the UFU is fully-depleted, the Android ‘Home’ screen
will appear after 40 seconds.
After another 30 seconds, the TruNode®
application will launch
and the UFU will be ready to detect the TruNode®
probe.
Note: If the TruNode app icon is pressed before it opens
automatically it will still launch OK, but the warning below
may appear. If so, click OK.
Caution: If the battery charge is below 50%, turn off the
UFU and charge it prior to use. Do not use the Charger in
the operating room.