3.
WARNINGS and CAUTIONS
General
Failure to thoroughly review and adhere to the
information contained in this User Manual may pose a
potential hazard to the patient and/or the user and may
void the warranty.
The TruNode
®
Gamma Probe System is not intended for
use in the central nervous system.
Do not modify this equipment without authorization of
the manufacturer
User Feedback Unit
Use only the Faxitron Power Supply provided with the
UFU for recharging. Use of another power supply may
increase the risk of electric shock
Fully charge the UFU before using the system.
Do not use the Charger in the operating room.
Do not use the Charger in the presence of an oxygen-
enriched environment or in the presence of a
flammable anesthetic mixture with air or with oxygen or
nitrous oxide.
Keep the UFU off when charging.
Remove the Charger when the UFU is fully charged.
The UFU and Charger are non-sterile. Do not sterilize
these components.
If liquid is spilled on the UFU, remove it from service and
contact Faxitron for instructions to return for inspection.
Ph: 1. 877.910.0030.
Securely mount the UFU to an IV pole.