MODEL
 
RANGE 
Switch S4
 
MIRROR
Selector
DT-906
 
61 m (200')
OPEN* 
200'*
DT-906
 
37 m (120')
CLOSED
 120' 
DT-900
 
27 m (90') 
OPEN* 
90' *
 
DT-900
 
15 m (50') 
CLOSED
 
50' 
SENSITIVITY 
S3 
HIGH
H
NORMAL
 
N*
 
LOW
 
L**
 
DT·900 Series 
DUAL TEC
® 
Motion Sensor for Commercial
 
and 
Light Industrial Applications - Installation Instructions
 
Step 1 
Select mounting height. 
 
Step 2 
Carefully push screwdriver into slots
 
to 
disengage latches and open top
 
cover. 
 
Step 3 
Firmly insert screwdriver into slot in  arrow 
and rotate PIR Mirror Selector
 
to the 
correct range. 
 
Step 4 
Set switch S4 to establish microwave
 
range. 
 
 
Step 5 
Locate correct sensor range scale
 
and 
rotate Vertical Adjustment Screw
 
until the 
diamond corresponds to the
 
sensor 
mounting height (coarse
 
adjust). 
 
 
 
 
 
 
 
 
 
NOTE:  Fine adjust may be needed during
 
walk-test. See Supplemental Information. 
Step 6 
Set switch S3 to establish the sensitivity 
best suited to your application. 
 
 
 
 
 
 
 
 
 
 
 
*Factory default setting.  
**Not connected 
**Not recommended for DT-906 
Step 7 
Select  INFORMER® 
 
mode with switch
 
S2 if desired. (See Supplemental
 
Information). 
Step 8 
Carefully push screwdriver into slot  to 
disengage latch and remove
 
bottom 
cover. 
 
Step 9 
Unfasten screws and remove mounting 
plate from sensor. 
Step 10 
Attach mounting plate to wall at
 
desired 
height, using four fasteners
 
(not 
supplied). 
 
Step 11 
Install M5 (#10) screw in wall 1.9 cm
 
(3/4") below mounting screw, as
 
shown, 
for tamper activation. 
1.9 cm (3/4") 
Step 12 
Pull about 30 cm (12") of wire from  wall 
through the opening in the
 
mounting 
plate and route wire to the
 
terminal strip. 
Step 13 
Hang the sensor on the mounting
 
plate 
hooks and fasten with the two  mounting 
plate screws. 
 
 
Step 14 
Wire the unit 
as shown.  
 
 
Use 2.0 - 0.3 mm
2 
(14 – 22 AWG) 
 
 
 
 
 
NOTE: Secure 
wires to 
mounting plate with  
tie wraps. 
 
Step 15 
Loosen horizontal locking screw in 
sensor support base. 
 
Step 16 
Grasp housing and rotate it to the
 
desired 
position (coarse adjust).  If
 
fine adjust is 
needed see Steps 20 -
 
22. 
NOTE: Reference marks 
= 
5° change.
 
Step 17 
Apply power to sensor and prepare
 
for 
walk-test. 
 
 Wait 90 seconds for power-up self-test 
to run.  All LEDs will flash. 
 
PIR  ALARM    MW
 
(Green)  (Red)    (Yellow)
 
 
 
 
 
 
 
 
NOTE
:  LEDs flashing after 90 sec. 
= 
defective
 
Step 18 
Turn the microwave potentiometer 
counterclockwise to decrease the
 
microwave range to minimum. 
 
During walk-test, gradually turn the
 
potentiometer clockwise increasing
 
microwave sensitivity until the
 
desired 
range is obtained. 
Step 19 
Walk-test the sensor to check for
 
adequate detection coverage and to
 
verify the sensor is fully functional.
 
Two to 
four normal steps should
 
make the LEDs 
light and trigger an  alarm. 
NOTE: If an on-going self-test problem, mask
 
condition or an INFORMER condition occurs, the 
LEDs display a pattern that identifies the
 
trouble. 
See Supplemental Information
 
(Table 3). 
 
NOTE:  When there is no motion in the
 
detection area, all three LEDs should be off. 
Step 20 
For finer horizontal adjustments,
 
loosen 
the PIR horizontal fine locking
 
screw on 
PCB. 
 
 
 
 
 
 
 
 
 
Step 21 
Rotate horizontal fine adjust knob to
 
the 
desired position. 
NOTE:  Fine adjustment allows for small
 
changes (3 degrees right or left) between
 
coarse settings. 
 
Step 22 
Tighten horizontal fine locking screw
 
on 
PCB. 
 
 
Step 23 
Tighten horizontal locking screw in 
sensor support base. 
Step 24 
Remove jumper at J5, on the PCB,
 
to 
disable the LEDs after walk-testing. 
 
 
 
 
 
 
 
 
 
 
Step 25 
Complete installation by closing top
 
cover and replacing bottom cover. 
 
 
 
 
DT‐906
DT‐900
 
30 cm (12”) minimum
4 m (12’)
2 m (6’)
Optimal