Taking the ERX Recorder “Un-Configured”
“Un-Configured” refers to resetting the instrument’s items to a firmware default.
Firmware defaults are different from the default values in the “Parameter Item List” section.
The default values listed in “Parameter Item List” are a combination of firmware and factory
defaults.
Once the ERX Recorder is “un-configured” Audit Trail, calibration informatio
n, and instrument
settings are all cleared. Therefore, it is important to record (down load an item file) this
information before taking the ERX Recorder un-configured. Specific information about
downloading an item file is available in MasterLink32 software.
There are two methods to put an ERX Recorder in un-configured state. The first is with
MasterLink32 Software, and the second is a hardwar
e method that is useful if a serial link cannot
be established.
Software method:
1. Make a physical connection to the ERX Recorder via serial link.
2. Start MasterLink32
3 Select form the Menu bar SetupAdvanceRaw Instrument AccessRaw Item
Access left click Raw Item Access