Honeywell ERX - Parameter Item List (P1 Calibration); P1 Calibration Zero Offset Settings; P1 Calibration Span Factor Settings

Honeywell ERX
84 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
45
The last recorded measurement of gas temperature and is scaled to the selected
temperature units at 551.
Default: -40.00
503 Case Temperature
The last recorded measurement of the air temperature inside the recorder, considered an
ambient temperature measurement.
Default: -40.00
504 Case Maximum Temperature
The highest recorded case temperature measurement since the recorder was last reset.
D
efault: -40.00
505 Case Minimum Temperature
The lowest recorded case temperature measurement since the recorder was last reset.
Default: 165.00
506 P1 Pressure Calibration Zero Offset
The most recent span value for P1 pressure transducer and obtained during a pressure
span calibration. A value that is normally very close to 1.0000 and recalculated on each
Pressure Span Calibration
.
Default = 0.000
507 P1 Previous Calibration Zero Offset
The previous P1 span factor recorded in item 506 and is updated on each calibration.
Default: 0.000
508 P1 Pressure Calibration Span Factor
The most recent span value factor for P1 pressure transducer and obtained during a
pressure span calibration. The value is normally very close to 1.0000 and recalculated on
each pressure span calibration.
D
efault = 1.0000
509 P1 Previous Calibration Span Factor
The previous P1 span factor recorded in item 508 and is updated on each calibration.
Default: 1.0000
510 P2 Calibration Zero Offset
The most recent span value for P2 pressure transducer and obtained during a pressure
span calibration. A value that is normally very close to 0.0000 and recalculated on each
Pressure Span Calibrat
ion
.
Default: 0.0000
511 P2 Previous Calibration Zero Offset
The previous P2 zero offset value recorded in item 510 and is updated on each
calibration.
Default: 0.0000
Item Item Name & Description
512 P2 Pressure Calibration Span Factor
The most recent span value factor for P2 pressure transducer and obtained during a
pressure span calibration. The value
is normally very close to 1.0000 and recalculated on
each pressure span calibration.