34-SM-03-01
Page 15
Zero Temperature Effect per 28°C
(50°F)
• Applies for model with Stainless Steel
barrier diaphragms
In Analog Mode: ±0.1125% of calibrated span.
For URV below reference point (300 psi), effect equals:
±0.0125 ± 0.10
()
300 psi
span psi
or ±0.0125 ± 0.10
()
21 bar
span bar
in % span.
In Digital Mode: ±0.10% of calibrated span.
For URV below reference point (300 psi), effect equals:
±0.10
()
300 psi
span psi
or ±0.10
()
21 bar
span bar
in % span.
Combined Zero and Span Temperature
Effect per 28°C (50°F)
• Applies for model with Stainless Steel
barrier diaphragms
In Analog Mode: ±0.25% of calibrated span.
For URV below reference point (300 psi), effect equals:
±0.15 ± 0.10
()
300 psi
span psi
or ±0.15 ± 0.10
()
21 bar
span bar
in % span.
In Digital Mode: ±0.225% of calibrated span.
For URV below reference point (300 psi), effect equals:
±0.125 ± 0.10
()
300 psi
span psi
or ±0.125 ± 0.10
()
21 bar
span bar
in % span.
Stability (At Reference Conditions)
±0.025% of URL per year.
Damping Time Constant
Adjustable from 0 to 32 seconds digital damping.