HPS Technical Support Page 3
Removing / Installing the Bezel assembly
Step
Action
ATTENTION!
Turn off the power to the analyzer.
More than one switch may be required to remove power.
1
Open the display board connector latch as shown below and remove Display
Cable from display board.
Also, slip the display cable out from under the plastic clip on the inside left of the
case.
2
Remove Phillips head screw in the center of the door hinge and remove the bar.
3
Hinge pins are now exposed. Swing them toward you, and then release the door by
sliding off the pins.
4
Slide the new Bezel Assembly (door) on the pins, replace bar and screw.
5
Insert the other end of the replacement display cable (foils should be facing
toward the display board) into the display board connector and close latch.
6
Close the Bezel and secure four captive screws to a torque value of .20Nm (1.5 Lb-in).
WARNING! Do not apply power until the bezel is closed.