DEVICE DESCRIPTION
2
11
VERSAFLOW VORTEX 200
www.honeywellprocess.com34-VF-25-137 iss.1 GLO July 2019 US
The permitted ambient temperature range is indicated on the nameplate; depending on the
device version it is T
amb
= -40...+65°C / -40...+149°F.
The minimum product temperature is -40°C/ -40° F.
When using in zone 21, observe the maximum permissible product and ambient temperatures
for temperature class 2 and the information for use of a heat-resistant connecting cable and
cable entry.
Max. permissible product and ambient temperatures with signal converter or
connection box mounted above the flow sensor
Figure 2-3: Signal converter above the flow sensor
Temperature class T6 T5 T4 T3 T2
T
amb
in °C 60 65 60 65 60 65 40 60 65 40 60 65
Nominal size
DN15…25 80 65 100 100 135 135 1 200 200 1 165 1 240 200 1 165 1
DN40…50 80 65 100 100 135 135 1 200 175 1 150 1 240 175 1 150 1
DN65…100 80 65 100 100 1 135 1 130 1 200 150 1 130 1 235 1 150 1 130 1
DN150...300 75 65 100 100 135 135 1 200 185 1 155 1 240 185 1 155 1
Table 2-7: Temperature class in °C
1 Permanent service temperature of connecting cable and cable entry min. 80°C
Temperature class T6 T5 T4 T3 T2
T
amb
in °F 140 149 140 149 140 149 104 140 149 104 140 149
Nominal size
DN15…25 176 149 212 212 275 275 1 392 392 1 329 1 464 392 1 329 1
DN40…50 176 149 212 212 275 275 1 392 347 1 302 1 464 347 1 302 1
DN65…100 176 149 212 212 1 275 1 266 1 392 302 1 266 1 455 1 302 1 266 1
DN150...300 167 149 212 212 275 275 1 392 365 1 311 1 464 365 1 311 1
Table 2-8: Temperature class in °F
1 Permanent service temperature of connecting cable and cable entry min. 176°F