Ambient CO monitor
APMA-370
Operation Manual [eng]
CODE:GZ0000051249J

Table of Contents

Questions and Answers

  • S
    Susan MeadowsAug 18, 2025
    Why is the output signal level too low on my horiba APMA-370?
    • C
      Christine FarmerAug 18, 2025
      The output signal level might be too low on your Horiba Measuring Instruments due to several reasons: * The range may not be set properly. You can fix this by setting the range correctly or selecting the automatic range. * The zero calibration might not be accurate enough. Perform the calibration again. * The span calibration might not be accurate enough. Perform the calibration again. * The catalyst may have deteriorated, in which case you should replace the catalyst pipe. * There might be pressure loss at the sample inlet or outlet. Check the piping for clogging and rectify it. * The filter could be contaminated, so replace it. The filter element should be replaced every 2 weeks.
  • M
    Michelle ChavezAug 20, 2025
    Why is the output signal level too high on my horiba APMA-370?
    • U
      uarroyoAug 20, 2025
      The output signal level might be too high because: * The range is not set properly. To fix this, set the range properly or select the automatic range. * The zero calibration is not accurate enough, so you should perform the calibration again. * The span calibration is also not accurate enough, so you should perform that calibration again as well.
  • P
    pam78Aug 23, 2025
    What to do if readings are not stable enough on horiba APMA-370 Measuring Instruments?
    • T
      Thomas WilliamsAug 23, 2025
      If the readings on your Horiba Measuring Instruments are not stable enough, it could be due to: * An unstable flow rate. Check the filter and replace it if it's contaminated. The filter element should be replaced every 2 weeks. * Check the pump and diaphragm. If they have been used exceeding their replacement time, replace them.
  • D
    Donna MejiaAug 26, 2025
    How to fix span calibration issues on horiba Measuring Instruments?
    • F
      Frank WilliamsAug 26, 2025
      Span calibration issues may arise in Horiba Measuring Instruments due to: * An unstable measured value. Perform the span calibration again. * An incorrect span gas concentration. Check the span gas concentration and perform the span calibration with gas of the proper concentration. * The span gas concentration being set improperly. Set the span gas concentration properly.
  • C
    Clifford JenkinsAug 28, 2025
    What to do if there is no output signal available on horiba APMA-370?
    • A
      amy32Aug 28, 2025
      If there is no output signal available, it may be due to: * A flow rate that is too low. Check the measured gas line and calibration gas line for disconnected piping. * A disconnected connector. Connect it properly.
  • D
    Diane FreemanAug 30, 2025
    How to resolve zero calibration issues on horiba APMA-370 Measuring Instruments?
    • A
      Alison JohnsonAug 30, 2025
      Zero calibration issues on your Horiba Measuring Instruments may occur because: * The measured value is not stable. Perform the zero calibration again. * Gas other than the zero gas has been supplied during the zero calibration. Supply the zero gas.
  • T
    Tina ThomasSep 2, 2025
    How to solve catalyzer temperature problems on horiba Measuring Instruments?
    • A
      Aaron MooreSep 2, 2025
      Catalyzer temperature issues on Horiba Measuring Instruments can be caused by: * The instrument not being sufficiently warmed up. Warm up the instrument. * The ambient temperature deviating from the specified range (0°C to 40°C). Allow the instrument to stand for a while in an environment of temperature within the specified ambient temperature range.
  • M
    mhughesSep 4, 2025
    How to address high noise level issues in horiba APMA-370?
    • W
      wademanuelSep 4, 2025
      A high noise level in Horiba Measuring Instruments can occur because: * Condensation occurs throughout the instrument. Allow the instrument to stand for a while in an environment of stable temperature within the operating temperature range until condensation disappears. * Significant vibrations occur in the installation site. Change the installation site.
  • S
    Stephen FarmerSep 7, 2025
    What causes BATT Battery error in horiba Measuring Instruments and how to fix it?
    • J
      Johnny MurphySep 7, 2025
      The BATT: Battery error in Horiba Measuring Instruments indicates that the battery has reached the end of its service life (about 3 years). Replace the battery. If the settings return to the default value after the battery has been replaced, set the current time and AIC start time again.
  • B
    Brenda CortezSep 9, 2025
    How to fix CF alarm issues on horiba APMA-370 Measuring Instruments?
    • V
      villaselenaSep 9, 2025
      The CF alarm on Horiba Measuring Instruments can be triggered by: * The CF not being connected. Insert a CF that has sufficient free space. * The CF having no free space. Format the CF or insert a CF that has sufficient free space.

horiba APMA-370 Specifications

General IconGeneral
Measurement MethodChemiluminescence
Linearity±1% of full scale
Resolution0.1 ppm
Sample Gas Flow Rate1 L/min
Power SupplyAC100V-240V, 50/60 Hz
Dimensions400 x 300 x 200 mm
Measurement PrincipleChemiluminescence
Sampling Flow Rate1 L/min
OutputDigital output (RS-232C)
Dimensions (Alt)400 x 300 x 200 mm
Weight (Alt)10 kg