Instruction Manual
Benchtop pH/ORP/ION/Conductivity Meter
LAQUA-PH2000
LAQUA-ION2000
LAQUA-EC2000
LAQUA-PC2000

Questions and Answers

  • A
    Angie EvansAug 19, 2025
    What to do if my Horiba Measuring Instruments shows DATA FULL?
    • J
      james01Aug 19, 2025
      If the Horiba Measuring Instruments displays 'DATA FULL', it means the maximum number of saved data has been exceeded. To resolve this, print or transfer all data, and then clear the data log.
  • L
    Lisa BriggsAug 21, 2025
    Why does the indicated value fluctuate on my Horiba Measuring Instruments when it is problem with the electrode?
    • S
      Sherry LeonardAug 21, 2025
      If the indicated value on your Horiba Measuring Instruments fluctuates, and it's due to the electrode, it could be for several reasons: * The electrode is dirty: Clean the electrode. * The electrode is cracked: Replace the electrode. * The wrong internal solution is being used: Use the correct internal solution. * There are air bubbles on the electrode: Shake the electrode to remove the air bubbles. * The level of internal solution in reference electrode is low: Replenish the internal solution of the reference electrode until it is higher than the level of the sample.
  • J
    John BrownAug 23, 2025
    Why does the indicated value fluctuate on my Horiba Measuring Instruments when it is problem with the instrument?
    • M
      Matthew RichmondAug 23, 2025
      If the indicated value on your Horiba Measuring Instruments fluctuates, and it's due to the instrument, it could be due to: * Electrical interference: Measure in a location free from induction and ground all AC-powered equipment. * Incorrect electrode connection: Ensure the electrode is properly connected.
  • N
    Nicole Olson MDAug 26, 2025
    Why is the repeatability of the measured value poor on my Horiba Measuring Instruments?
    • C
      Claudia JohnsonAug 26, 2025
      If the repeatability of the measured value is poor on your Horiba Measuring Instruments, it could be due to: * The effect of the sample solution: Repeatability becomes poor when the pH of the sample changes over time. * The electrode is dirty: Clean the electrode. * The electrode is broken: Replace the electrode. * The internal solution of the electrode runs out or contaminated: Replace the internal solution with a new one. * The level of internal solution in reference electrode is low: Replenish the internal solution of the reference electrode until it is higher than the level of the sample.
  • S
    Shawn StevensonAug 27, 2025
    What does NOT STBLE mean on my Horiba Measuring Instruments and how to fix it?
    • M
      marquezshannonAug 27, 2025
      If your Horiba Measuring Instruments shows 'NOT STBLE', it means the ENT key was pressed before the smiley icon and calibration value stabilized. Wait for the smiley icon and calibration value to stabilize before pressing the ENT key.
  • J
    jill54Aug 30, 2025
    How to fix OFFS ERR on Horiba Measuring Instruments?
    • A
      Anthony SimpsonAug 30, 2025
      If your Horiba Measuring Instruments displays 'OFFS ERR', it indicates that the electrode is either dirty or the reference junction is clogged. Clean the electrode to resolve this issue.
  • G
    Gabriel WilliamsAug 31, 2025
    What causes SLPE ERR on Horiba Measuring Instruments?
    • S
      Sara BrownSep 1, 2025
      If your Horiba Measuring Instruments displays 'SLPE ERR', it indicates that the electrode sensitivity is low. Clean and recalibrate with a fresh standard solution. If the problem continues, replace the electrode.
  • M
    Melanie MorrisSep 3, 2025
    What does BUFF ERR mean on Horiba Measuring Instruments?
    • J
      Jacob MurphySep 3, 2025
      If your Horiba Measuring Instruments displays 'BUFF ERR', it means the instrument cannot identify the pH buffer. Check the calibration solution and use a fresh one if necessary.
  • R
    Robert GallagherSep 5, 2025
    Why does the indicated value not change on my Horiba Measuring Instruments?
    • M
      millernicoleSep 5, 2025
      If the indicated value on your Horiba Measuring Instruments does not change, it could be due to several reasons: * The electrode is cracked: Replace the electrode. * The electrode is not connected correctly: Connect the electrode correctly. * The instrument is in HOLD state: Cancel the HOLD state. * Instrument defect: Consult your dealer.
  • S
    shermanthomasSep 7, 2025
    What does CAL DUE mean on my Horiba Measuring Instruments?
    • C
      Carolyn BarberSep 7, 2025
      If your Horiba Measuring Instruments displays 'CAL DUE' and the electrode icon blinks in MEAS mode, it means calibration was not performed at the designated interval. Perform calibration.

Summary

Preface

Warranty and responsibility

Details warranty terms and user responsibilities for the product.

For Your Safety

Hazard classification and warning symbols

Explains hazard levels (DANGER, WARNING, CAUTION) and warning symbols.

Safety precautions

Provides critical safety measures for instrument and electrode handling.

Product Handling Information

Operational Precautions (instrument)

Guidelines for safe and correct operation of the instrument.

Product Overview

Basic Operations

Turning On the Instrument

Step-by-step guide to power on the benchtop meter.

Connecting an Electrode

Procedure for properly connecting electrodes to the instrument.

Calibration

pH Calibration

Detailed steps for calibrating the pH measurement function.

ORP;mV Calibration

Procedure for calibrating the ORP/mV measurement.

Ion Calibration

Steps for calibrating ion concentration measurements.

Conductivity Calibration

Guide for calibrating electrical conductivity measurements.

TDS Calibration

Information on TDS calibration, noting it's derived from conductivity.

Salinity Calibration

Procedure for calibrating salinity measurements.

Temperature Calibration

Steps to calibrate the temperature sensor for accurate readings.

Data

Setup

P1 pH Setup

Configuration options for pH measurement and calibration settings.

P1 Ion Setup

Configuration options for ion measurement and calibration.

P1 COND Setup

Configuration options for conductivity measurement and calibration.

P2 TDS Setup

Configuration options for Total Dissolved Solids (TDS) measurement.

P3 SAL Setup

Configuration options for salinity measurement.

Data Setup

Configuration options for managing data logging.

General Setup

Configuration options for general instrument settings.

Clock Setup

Procedure for setting the instrument's real-time clock.

Maintenance and Storage

Maintenance and Storage of the Instrument

Guidelines for cleaning and storing the main instrument.

Maintenance and Storage of Electrodes

Overview of maintaining and storing various types of electrodes.

Error Messages and Troubleshooting

Error Messages

Lists common error codes, their definitions, causes, and solutions.

Troubleshooting

Provides solutions for common problems like fluctuating values or connectivity issues.

Meter Specifications

pH Range

Specifies the measurement range and resolution for pH.

horiba LAQUA-PH2000 Specifications

General IconGeneral
pH Resolution0.01 pH
pH Accuracy±0.01 pH
mV Resolution1 mV
Temperature Resolution0.1 °C
Temperature Accuracy±0.5 °C
Calibration PointsUp to 5 points
ConnectivityUSB
WaterproofIP67
Temperature CompensationAutomatic
DisplayLCD