Model  333A/334A 
Section 
V 
SECTION 
V 
MAINTENANCE 
5-1. 
I 
NTR ODUCTI 
0 
N. 
5-2.  This section contains maintenance and service 
information for the 333A and 334A Distortion 
Analy- 
zers.  Included are Performance Checks, Adjustment 
and  Calibration 
Procedures, 
and Troubleshooting 
Techniques. 
5-3. 
REQUIRED 
TEST 
EQUIPMENT. 
5-4.  The 
e 
uipment needed to properly maintain the 
Model 333A7334A 
is 
listed  in Table 5-1. 
If 
the  re- 
commended  model 
is 
not 
available,  any  instrument 
that has specifications equal to 
or 
better than the 
re- 
quired specifications may 
be 
used. 
5-5. 
PERFORMANCE  CHECKS. 
5-6. 
The performance checks 
are 
in-cabinet  proce- 
dures that can 
be 
used to verify instrument perform- 
ance.  These procedures  can 
be 
used for periodic 
maintenance,  to check  specifications after a repair, 
or 
for 
incoming quality control inspection.  Perform- 
ance checks 
for 
both the 333A and 334A are provided. 
The performance  checks 
are 
applicable toboth instru- 
ments except where noted 
in 
paragraph heading. 
5-7.  The performance checks are made with the ac 
power  cord connected to nominal line voltage 
(lOOv/ 
120V/ 220V/  240V,  48 to 
66 
Hz), 
MODE 
SWPTCH 
to 
MANUAL, 
HIGH 
PASS FILTER to 
OUT, 
and 
NORM. 
- 
R 
F.  DET.  switch to 
NORM, 
unless otherwise 
specified. 
5-8. 
A 
Performance Check Test Card 
is 
provided 
at 
the end of  this section for recording the performance 
of 
the instrument 
during 
the performance checks.  The 
card may 
be 
removed from the manual and used 
as 
a 
permanent  record 
of 
the  incoming  inspection 
or 
of 
a 
routine performance check. 
If 
the instrument 
fails 
to meet any 
of 
its 
specifications,  perform theAdjuat- 
ment and Calibration Procedures,  Paragraph 5-23. 
5-9. 
VOLTMETER ACCURACY 
AND 
FREQUENCY 
RESPONSE CHECK. 
a. 
b. 
C. 
d. 
e. 
f. 
g. 
Conned voltmeter  calibrator to 333A/334A 
as 
shown 
in 
Figure 5-1.  Test 
oscillator 
will 
be 
used later 
in 
check. 
METER. 
Set 
voltmeter  calibrator  for  400  Hz rms 
output. 
Check 333A/334A  voltmeter 
full 
scale read- 
ing 
on 
all ranges 
against 
appropriate 
rms 
input 
from 
voltmeter calibrator.  Voltmeter 
accuracy should be within 
&2%. 
Disconnect voltmeter calibrator. 
Set 333A/334A  FUNCTION Switch to VOLT- 
Connect 
5052 
output 
of 
test oscillator,  term- 
inated in 5052, 
to 
333A/334A. 
Set 333A/334A  METER RANGE  switch 
to 3 
VOLTS. 
VOLTMETER  CALIBRATOR 
hp 
738BR  DISTORTION  ANALYZER 
hp 
333A/334A 
c. 
r~ 
tt 
00 
0 
(9 
' 
' 
006 
I 
I 
I 
I 
I 
I 
TEST  OSCILLATOR 
I 
hp 
6521 
I 
- 
son 
FEED-THRU 
I 
TERMINATION 
I 
I 
I 
I 
------------ 
SSA-B-1468 
9-8 
fiL 
Figure 5-1. 
Voltmeter 
Accuracy and Frequency Response Check 
5-1