Model 333A/334A 
SENSITIVITY 
Switch 
Position 
4 
Position 
3 
Position 
2 
Position 
1 
C. 
d. 
e. 
f. 
g* 
h. 
TEST OSCILLATOR 
ADJUST 
OUTPUT (Increase)  FOR 
9 
mV 
+lo 
dB 
ac4 
+20 
dB 
s2c3 
+30 
dB 
s2c2 
+40 dB 
ac1 
SENSITIVITYVERNlER.. 
..... 
MIN. 
SENSITIVITY. 
.. 
.1 
step counterclochise 
METER RANGE 
. 
, 
. 
. 
, 
. . 
SET LEVEL 
MODE.. 
............. 
MANUAL 
FREQUENCY RANGE 
........ 
X1K 
Frequency 
Dial. 
............ 
60 
Turn slot 
on 
trimmer capacitor C3 parallel 
with side casting of  instrument. 
Set test oscillator frequency to 
60 
kHz, 
and 
adjust amplitude for  indication 
of 
+2 
dB 
on 
333A/334A  meter. 
Set 333A/334A  COARSE  and 
FINE 
BALANCE 
controls to center of  range. 
Set  333A/334A  FUNCTION  switch to 
DIS- 
TORTION. 
Adjust 
test 
oscillator frequency vernier con- 
trol and  C3 for  definite  null  indication  on 
meter  of  333A/334A.  Reduce  METER 
RANGE  setting 
as 
necessary to maintain 
on-scale  reading. 
Observe  electronic  counter  indication. 
If 
test 
oscillator  frequency 
is 
not  between 
57 
kHz 
and 
63 
kHz, 
333A/334A  frequency 
dial 
is 
out 
of 
tolerance.  Follow  steps 
be- 
low to bring frequency 
dial 
within tolerance: 
Set 333A/334A controls 
as 
in 
step b. 
Set test oscillator frequency for elec- 
tronic counter indication 
of 
60.00 
HI, 
and adjust amplitude for indication of 
+2 
dB 
on meter 
of 
333A/334A. 
Set 333A/334A  COARSE  and 
FINE 
BAL- 
ANCE  controls to center of  range. 
Adjust 333A/334A frequency 
dial 
vernier 
control for 
definite 
null. 
Remove knob from frequency 
dial. 
Do 
not change tuning capacitor, 
C4, 
setting 
Loosen retaining screws in frequency 
dial 
plate,  and slip 
dial 
until 60 
is 
dir- 
ectly under 
mark 
on indicator. 
Be 
careful not to change 
C4 
setting. 
Tighten retaining screws and replace 
knob. 
5-30.  SENSITIVITY SWITCH  CALIBRATION. 
a. 
Connect test oscillator to 333A/334A 
ae 
shown 
in 
Figure 5-1. 
NOTE 
Selector positions for the SENSI- 
TMTY selector 
will 
be 
referred to 
as 
follows: 
MIN 
= 
Position 
1, 
next 
step 
= 
Position 
2, 
etc.,  to MAX 
= 
Position 
6. 
b. 
C. 
d. 
e. 
f. 
8. 
Section 
V 
Set 
Distortion Analyzer controls 
as 
follows: 
FUNCTION.. 
........ 
SETLEVEL 
METER RANGE 
....... 
0.01 
VOLTS 
SENSITIVITY 
.......... 
Position 5 
SENSITIVITY VERNIER 
....... 
MAX. 
Set test oscillator frequency 
to 
400 
Hz, 
ad- 
just  amplitude 
for 
indication  of 
9 
mV  on 
333A/334A 
meter. 
Set 
a 
reference on meter 
of 
test  oscillator 
and 
use 
amplitude  control to maintain 
set 
reference whenever frequency of  oscillator 
is 
varied. 
Set test oscillator frequency to 100 
kHz. 
Adjust S2C5  for 333A/334A  meter indication 
of 
9 
mV. 
Set 333A/334A  SENSITIVITY switch to each 
position listed in Table 5-6,  and  increase 
test oscillator 
output 
by 
10 
dB 
for 
each re- 
spective  decrease 
in 
SENSITIVlTY switch 
position.  Adjust appropriate capacitor for 
333A/334A 
meter indication 
of 
9 
mV 
at 
each 
setting. 
NOTE 
When switching the SENSITMTY 
positions, 
be 
careful not to move 
SENSITIVITY VERNIER  control 
from 
MAX. 
position. 
5-31.  ADJUSTMENT 
OF 
FACTORY SELECTED 
COMPONENTS. 
5-32.  Certain components in the 
Model 
333A/334A 
are 
individually  selected to compensate for 
va.rying 
circuit parameters.  These components are designated 
lqr 
an 
asterisk 
(*) 
on the  schematic and in the 
parts 
list, 
and the typical 
value 
is 
shown 
describes the function of  the factory selected compo- 
nents 
and lists their value 
ranges. 
Table 
5-7 
5-33, 
TRO 
U 
BLES 
H 
0 
OTI 
N 
G 
PR 
0 
CEDU 
R 
Et. 
5-34.  This section contains procedures designed to 
assist 
in the  isolation of  malfunctions,  These 
pro- 
cedures 
are 
based  on 
a 
systematic 
analysis 
of 
the 
instrument circuits in 
an 
effort to localize the prob- 
Table 
5-6. 
Sensitivity Switch Calibration 
5-11