Chapter 4 Retrofitting Options
To Retrofit A6 High Stability Timebase Assembly (Options 001, 010, and
012)
Assembly-Level Service Guide 4-13
4
Figure 4-2. High Stability Timebase Retrofitting (Option 001, 010, 012)
10 Re-install the cover by performing the cover removal procedure,
in Chapter 3 of this guide, in reverse.
This completes A6 High Stability Timebase Assembly retrofit procedure.
11 Perform the appropriate timebase calibration procedure as found
on page 2-23 and page 2-24.