Chapter 1 Performance Tests
Test 6: Option 030/050/124 Channel 3 Frequency Accuracy
(HP 53131A/132A Only)
Assembly-Level Service Guide 1-35
1
Test 6: Option 030/050/124 Channel 3
Frequency Accuracy (HP 53131A/132A Only)
This test verifies the frequency accuracy of the optional 3-GHz or 5-GHz
Channel 3 (Option 030).
Equipment
HP 8663A Synthesized Signal Generator (Option 030 only)
HP 8340B Sweep Oscillator
HP 1250-0780 Type N-to-BNC Connector (Option 030 only)
HP 10503A BNC Cable (2)
HP 10100C 50
Feedthrough
HP 11500D N-N Cable (Option 050 only)
HP 1250-1250 N(M) to SMA(F)
HP 1250-0777 N(F) to N(F)
Counter Setup
1Cycle the POWER key to preset the Counter.
2 Set the Counter as follows:
Freq & Ratio,
FREQUENCY 3
Gate & ExtArm,
TIME: TIME 1.000 s
Procedure
1 Connect equipment as shown in Figure 1-9.
(NOTE: For Options 050 and 124, skip steps 2 through 5.
Proceed to step 6.)
2 Set the HP 8663A Synthesized Signal Generator to output
100 MHz at 17.0 dBm.