EasyManua.ls Logo

HP 61 - Calibration Procedure

HP 61
9 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appl.
Note
61
For
coaxial
testing
use
the
%
780
series
Directional
Detectors*
(shown
in
Figure
2)
to
complete
the
level-
ing-feedback-loop
from
RF
to
the
sweep
oscillator
leveling
amplifier.
Frequency
response
of
these
de-
vices
for
1
to
2
and
2
to
4
Gc
ranges
is
+0.2
db.
All %
752
series
waveguide
couplers
have
a
RF
fre-
quency
response
of
better
than
£0.5
db.
To
minimize
this
frequency
error
in
making
transmission
meas-
urements
use
the
coupler
curve
matching
setups
shown
in
Figures
3
and
4.
Figure
3
shows
how
the
detected
RF
response
of
one
coupler
is
used
to
adjust
sweep
oscillator
leveling.
This
compensates
for
the
detected
RF
response
of
the
coupler
driving
the
oscil-
loscope,
if
the
two
sets
of
coupler
curves
are
identical.
Frequency
response
curves
of
752C
couplers
usually
match
within
one
or
two
tenths
of
a
db.
The
752C/
424A
system
used
as
a
detector
presents
a
main
guide
SWRof
1.05to
the
device
under
test
and
therefore
re-
duces
measurement
mismatch
error.
The
424A
De-
tectors
have
a
response
of
+0.2db
through
H
band
and
+0.3
db
in
X
band.
The
response
of
M
and
P
band
units
is
0.5
db.
For
making
high-loss
transmission
measurements,
Figure
4
shows
an
arrangement
for
removing
coupler
attenuation
between
the
device
under
test
and
the
detec-
tor-indicator.
The
result
is
a
10-db
improvement
in
sensitivity over
the
Figure
3
setup.
This
system
also
employs
coupler
frequency
response
curve
compensa-
tion,
but
needs
the
factory
matched
E01-752C
arrange-
ment
in
order
to
hold
coupler
matching
limits
within
one
or
two
tenths
of
a
db.
Knowing
your
measurement
frequency
range,
select
the
proper
sweep
oscillator,
directional
coupler
or
directional
detector,
and
other
instruments
from
Table
IV.
For
the
oscilloscope
detectors
shown
in
Figures
2,
3,
or
4,
always
use
the
load
resistor
purchased
under
423A
or
424A
Option
02,
because
detector
and
load
have
been
optimized
for
square-law
voltage
response
(linear
power).
The
423A
or
424A
used
with
its
indi-
vidually
factory-mated
load
has
a
square-law
response
within
+1/2
db
below
50
mv
DC
output
as
measured
on
the
oscilloscope.
When
square-law
loading
is
re-
quired
for
the
detector,
use
only
the
factory-supplied
load
furnished
with
each
Option
02
detector.
Identifi-
cation
is
easy
because
load
and
detector
bear
identical
serial
numbers.
The
detector
used
for
leveling
need
not
be
square-law
loaded
because
under
closed
loop
conditions,
level
always
remains
the
same.
CALIBRATION
PROCEDURE
With
the
instruments
set
up
as
shown
in
Figures
2,
3,
or
4,
connect
coax
or
waveguide
together,
in
place
of
device
under
test.
Oscilloscope
horizontal
and
ver-
tical
amplifiers
should
be
set
to
respond
to
DC,
and
the
vertical
attenuator
should
be
set
to
a
convenient
multiple
or
fraction
of
ten.
Of
course,
the
maximum
power
applied
to
the
flat
detector
should
be
within
the
square-law
range.
Filters,
pads,
and
adapters
with-
in
the
RF
leveling
loop
will
not
affect
the
frequency
response
of
the
leveled
RF
output.
*
The directional
detector
combines
directional
coup-
ler
and
microwave
crystal
detector
thereby
avoiding
the
SWR
ambiguity
of
connecting
the
two
instruments.
Page
3
Transmission
Attenuation
in
db
1.4 1.5 1.6
Frequency
in
Gc
Figure
5.
Shows
0, 3, 7,
and
10
db
Calibration
Lines
and
the
Insertion-Attenuation
of
a
1200-Mc
Low-Pass
Filter
With
Sweep
Oscillator
Power
Switch
in
"STANDBY''.
Cut
transparent
""db"
scale
from
foil.
Place
scale
on
oscilloscope
graticule
as
shown
in
Figure5.
Use
either
static
charge
or
several small
spots
of
rubber
cement
to
hold
scale
in
place.
Adjust
sweep
oscillator
frequency
limits.
Set
oscilloscope
horizontal
gain
control
to
give
de-
sired
calibration
in
Ge/em.
The
sweep
oscillator
sweep
circuits
must
be
operating
for
this
adjustment.
Set
oscilloscope
vertical
position
control
to
estab-
lish
the
base
line
on
the
CRT
screen
for
«
attenua-
tion.
For
use
with
the ""db"
scale
in
Table
III,
the
base
line
should
be
exactly
3
cm
up
from
the
CRT
graticule
center
line.
With
Sweep
Oscillator
Power
Switch
in
"RF"'.
-
Adjust
sweep
oscillator
for
leveled
sweep
as
dis-
played
on
oscilloscope.
Check
flat
detector
for
overloading.
Detector
out-
put
as
measured
on
oscilloscope
should
not
exceed
50
mv.
If
voltage
is
excessive,
lower
sweep
oscil-
lator
output
power
or
insert
an
appropriate
pad be-
tween
RF
and
detector.
Set
zero
db
calibration
line
exactly
3
cm
down from
CRT
graticule
center
line,
using
oscilloscope
ver-
tical
attenuator
and
vernier.
Recheck
zero
and
«
dblines
again
as
outlinedunder
the
previous
steps.
Also,
be
sure
oscilloscope
ver-
tical
DC
balance
is in
proper
adjustment
sothe
base
line
calibration
will
not
change
with
various
oscillo-
scope
attenuator
settings.

Related product manuals