EasyManua.ls Logo

HP 61 - Reflection Measurements

HP 61
9 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page
4
-
Using
sweep
oscillator
marker
controls,
check
lin-
earity
of
your
horizontal
sweep
at
major
vertical
graticule
lines
and
make
any
horizontal
adjustments
needed.
YOUR
AMPLITUDE
AND
FREQUENCY
SCALES
ARE
NOW
CALIBRATED.
MEASUREMENTS
Place
the
unknown
device
under
test
between
the di-
rectional
coupler
and
the
flat
detector
(feeding
the
os-
cilloscope)
shown
in
Figures
2,
3,
or
4.
Using
the
db
scale
shown
in
Table
III,
the
oscilloscope
presen-
tation
is
now
a
direct
plot
of
transmission
loss
in
db
vs
frequency.
Use
the
oscilloscope
vertical
step-
attenuator
to
bring
the
trace
as
close
as
possible
to
(but
still
above)
the
lower
3-cm
calibration
line,
with-
out
changing
the
vernier
or
vertical
position controls.
Note
the
ratio
of
steps
traversed
onthe
attenuator,
and
from
Table
I
add
the
related
number
of
db
to
the
cali-
brated
oscilloscope
scale
reading.
Table
I.
Oscilloscope
Scale
Correction
with
Attenuator
Setting
Attenuator
Voltage
Ratio
DB
Correction*
1.0
0
.5
3
.2
7
pat
10
.05
13
.02
17
.01
20
*
Add
this
number
to
transmission
attenuation.
Attenuator
settings,
of
course,
can
be
extended
be-
yond
those
shown
in
Table
I
using
the
same
system
of
correction.
If
more
vertical
amplifier
gain
is
needed,
use
the
140A
Oscilloscope
with
1400A
and
1420A
Plug-
In
Units.
The
photo
shown
in
Figure
5
is
an
oscillo-
scope
picture
of
a
1200-Mc
coaxial
low
-
pass
filter
response complete
with
vertical
attenuator
calibration
lines.
The
setup
of
Figure
2
was
used
for
this
picture;
arrows
show
sweep
oscillator
marker
pips
for
exact
calibration.
Ahigh
degree
ofaccuracy
can
be
attained
on
bothfrequency
and
amplitude
axes.
Note
the
trace-
splitting
precision
with
which
amplitude
and
frequency
calibration
lines
fall
on
the
graticule
in
the
example
given.
It
is
reasonable
to
expect
similar
performance
in
day-
to-day
service
with
truly
square
-
law/frequency
-
flat
detectors,
with
the
sweep
oscillator
well
leveled,
with
a
modern
internal
-graticule
oscilloscope,
and
with
stable
circuitry
throughout
the
system.
The
@ instru-
ments shown
in
Table
IV,
when
connected
as
shown
in
Figures
2,
3,
or
4,
incorporate
all
these
features.
REFLECTION
MEASUREMENTS
Probably
one
of
the
most
useful
functions
of
a
leveled
sweep
oscillator
is
in
making
rapid
accurate
reflection
measurements.
Typical
leveled
reflectometer
setups
for
coaxial
and
waveguide
testing
are
shown
in
Figures
6
and
7.
The
forward-detected
signal
is
used
to
level
Appl.
Note
61
the
sweep
oscillator.
The
detected
signal
reflected
from
the
device
under
test
is
displayed
on
the
oscil-
loscope
which
is
calibrated
in
SWR
vs
frequency.
The
unknown
load
on
which
the
reflection
measurements
are
being
made
is
used
as
atermination
on
the
through
arm
of
the
coupler
system,
opposite
the
end
fed
by
the
leveled
sweep
oscillator.
For
coaxial
testing,
use
the
&
770
series
Dual
Direc-
tional
Couplers
shown
in
Figure
6.
Coupler
arm
fre-
quency
response
is
specified
not
more
than
+1
db.
However,
if
the
response
of
forward
and
reverse
arms
match,
the
sweep
oscillator
will
be
leveled
to
the
for-
ward arm
response
curve
in a
manner
that will
com-
pensate
for
reverse-arm
response.
Typical
forward-
and
reverse
-
arm frequency
-
response matching
for
these
devices
in
the
1
to
2and
2
to
4
Ge
range
is
better
than
0.2
db.
For
waveguide
testing,
two 752
directional
couplers
can
be
placed
together
to
form
a
waveguide
reflectom-
eter.
Waveguide
coupler
matching,
as
in
dual
coaxial-
coupler
arm-matching,
uses
the
forward-arm
response
to
compensate
for
reverse-arm
response.
Coupler-
arm
frequency
-
response matching
for the
752C
ar-
rangement
shown
in
Figure
7
is
usually
within
one
or
two
tenths
of
a
db.
Knowing
your
measurement
fre-
quency
range,
select
the
proper
sweep
oscillator
and
directional
coupler
from
Table
IV.
Always
use
the
square-law
load
resistor
purchased
under
Option
02
for
loading
423A
or
424A
detectors
as
indicated
in
the
previous
section
on
transmission
measurements.
Detector
DC
output
should
be
held
below
50
mv.
CALIBRATION
PROCEDURE
With
the
instrument
set
up
as
shown
in
either
Figure
6
or
7,
oscilloscope
horizontal
and
vertical
amplifiers
should
be
set
to
respond
to
DC,
and
the
vertical
atten-
uator
should
be
set
to
a
convenient
multiple
or
frac-
tion
of
ten.
Power
applied
to
the
flat
detector
driving
the
oscilloscope
shouldbe
within
the
square-law
range.
Filters,
pads,
and
adapters
within
the
RF
leveling
loop
will
not
affect
the
frequency
response
of
the
leveled
RF
output.
Be
sure
reflectometer
coupler
directivity
is
high
enough
so
that
unwanted
forward
signals
do
not
inter-
fere
with
reflected
signals.
Table
II
shows
the
mini-
mum
quality
of
coupler
directivity
requiredfor
meas-
uring
reflections.
The
higher
the
coupler
directivity
above
the
minimum
value,
the
lower
the
area
of
meas-
urement
ambiguity.
Table
II.
Coupler
Directivity
Required
for
Measuring
a
Given
SWR
SWR
Measured
|
Coupler
Directivity
Required
>1.10
>30
>1.05
>35
>1.03
>40

Related product manuals