Procedure: Changing the Lit offset setpoint
Lit offset
ąCONFIGURE FRONT DET
Mkup gas type N2
Lit offset 2.0 <
Electrometer On
1. Press [Config][Front Det] or [Config][Back Det].
2. Scroll to
Lit offset and enter a number.
The default is 2.0 pA.
Enter 0 to disable the automatic reignite function.
The setpoint range is 0 to 99.9 pA.