EasyManua.ls Logo

HP StreamSmart 400

HP StreamSmart 400
60 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Advanced Topics
34
h
3. Figure 4-12 shows the calibration screen, with its options for 1- or 2-point
calibration. Note that the meter continues to show the live streaming of
sensor values. Press %1$ to enter the correction for the current reading. The
current reading is still 6.408, but it should read -6.433. Enter the correct
value as shown in Figure 4-13 and press $OK# to return to the calibration
screen.
4. Note that the calibration change is not yet effective, as you have the option
of entering a second correction. The white square in the %$1!!qqq!!! key in
Figure 4-14 indicates that the first point correction has been logged.
5. Press $2$ to enter a second point correction, or press $OK# to make the 1-
point calibration effective (Figure 4-15). Of course, you can press
CANCL to
abort the calibration process.
Figure 4-10 Figure 4-11
Figure 4-12 Figure 4-13
Figure 4-14 Figure 4-15

Related product manuals