EasyManuals Logo

HT JUPITER User Manual

HT JUPITER
44 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #18 background imageLoading...
Page #18 background image
JUPITER
IT - 16
4. In conditions of correct voltage and frequency, the instrument shows message HOLD,
symbols and PH1and the buzzer sounds continuously, waiting for a stable voltage
value on phase L1 to be detected (see Fig. 14 left side).
Fig. 14: Detecting phase L1 and waiting for phase L2
5. Do not remove the lead from phase L1 until message PH 2appears flashing on the
display (see Fig. 14 right side)
Fig. 15: Detecting phase L1 and waiting for phase L2
6. Position the red lead onto phase L2 of the three-phase system to be tested (see Fig.
12). In case passage between phase L1 and phase L2 takes more than 10s, the
instrument shows message t.outon the display (see Fig. 15 left side). In conditions
of correct voltage and frequency, the instrument shows message HOLD, symbols
and
PH2and the buzzer sounds continuously, waiting for a stable voltage value on
phase L2 to be detected (see Fig. 15
right side)
7. Upon detection of a stable voltage value on phase L2, the instrument automatically
shows message 1.2.3.(test OK) or message 2.1.3(test NOT OK) as shown in Fig.
16.
Fig. 16: Results of phase sequence and phase concordance test
8. In case it is necessary to check phase concordance between two parallel three-phase
systems, after detection of phase L1 of the first system, position the lead on phase L1
of the second system. The correct final result is message 1.1-(see Fig. 16 right side)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HT JUPITER and is the answer not in the manual?

HT JUPITER Specifications

General IconGeneral
BrandHT
ModelJUPITER
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals