EasyManua.ls Logo

HT Neptune - Page 30

Default Icon
46 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
NEPTUNE
EN - 28
measurement and comparing this new value to the compensated one. If the difference
between the values is 0.01, calibration is confirmed and the message OK is
displayed (see Fig. 33 middle). Subsequently, the instrument goes back to the
measuring screen with the message ZERO displayed to indicate that cable
compensation is present.
5. In case the first measurement detects a resistance of test cables >5.00, the
instrument shows the following screens in a quick sequence:
Fig. 34: Results of zeroing not correctly performed
6. Messages >5.00and NOT OKare initially shown on the display (see Fig. 34 left
side). Subsequently, message CLr is shown, to indicate that calibration has been
cancelled (see Fig. 34 middle) and the message ZERO is not shown in the
measuring screen (see Fig. 34 right side).
7. To zero a cable calibration in the instrument, carry out the procedure with open input
terminals V /+ and COM/ and press and hold (>2s) key /ZERO. The following
screens are displayed in a quick sequence:
Fig. 35: Zeroing calibration

Table of Contents

Related product manuals