Ending CC Temperature control
The temperature control can be ended at any time by pressing the function key T3 (65).
The temperature control and circulation is immediately stopped. Alternatively, to set the
unit to standby mode and use the main menu point Start & Stop - menu point Stop
temperature control.
For units with compressor:
The main switch (36) can be used to turn the power off only after the compressor has
been stopped by the controller.
Room temperature should be reached before the temperature control is ended.
Protect your application by NOT closing any isolation and drain valves. Overpressure /
underpressure can be built up within the system if valves are closed and temperature
adjustment takes place.