Speci cations
5. Speci cations
5.1 Equipment Classi cation
Type of protection
against electric shock.
Internally powered equipment
Degree of protection
against electric shock
Type BF - equipment with an applied part.
Mode of operation. Continuous
Degree of protection
against harmful ingress
of particles and/or water.
Main Unit: IP20
FD1 / FD3 probes : IPX7
All other probes (Tip only): IPX1
Degree of safety of
application in the
presence of a ammable
anaesthetic
Equipment not suitable for use in the presence of
a FLAMMABLE ANAESTHETIC MIXTURE WITH
AIR, OXYGEN OR NITROUS OXIDE
5.2 Standards Compliance
IEC60601-1: 1988 + A1:1991 +A2:1995
UL60601-1 : 2006
CSA C22.2 No 601.1-M90 (R2005)
5.3 FHR Performance*
Standard Mode Range - 60-210bpm Resolution - 1bpm
Averaging - 4 beats Accuracy - ±3bpm
Smoothed Mode
(FD2 only)
Range - 60-210bpm Resolution - 1bpm
Averaging - 8 beats Accuracy - ±3bpm
Manual Mode
(FD2 only)
Range - 60-210bpm Resolution - 1bpm
Averaging - 10 beats Accuracy - ±3bpm
*(excluding user error)