4. Unpacking and Installation
4.1. Unpacking
IMPORTANT NOTICE
PERSONNEL QUALIFICATIONS
Only qualified personnel is permitted to unpack, install and operate the KermaX-
plus
®
.
IMPORTANT NOTICE
TRANSPORTATION
During transportation the KermaX-plus
®
should be adequately protected in the
originally supplied or equivalent packing.
IMPORTANT NOTICE
DOCUMENTATION
The documentation supplied with the KermaX-plus
®
consists of this manual and
the factory calibration certificate.
4.2. Installation
4.2.1. Installing the Chamber in the X-Ray Machine
Slide the chamber into the lower accessory rails of the collimator. Ensure that even if the collimator
shutter is fully open that no shadows from the chamber frame are visible in the light field.
For chambers integrated into the X-ray collimator, please refer to the respective instruction manual from
the collimator manufacturer.
CAUTIONS
SAFETY PRECAUTIONS
When installing the device, please observe the safety precautions described in
Section 2.5 Safety Precautions.
CAUTION
TURN OFF THE POWER BEFORE CONNECT/DISCONNECT THE CHAMBER
Do not connect/disconnect the chamber to/from the X-ray machine when the
power is turned on. It may damage the electronics.