8
Icare ic100 tonometer manual
Atmospheric pressure: 500 hPa-1,060 hPa
Mode of operation: continuous
PERFORMANCE DATA
The performance data is obtained from
a clinical study, performed according to
American National Standard ANSI Z80.10-
2003 and International Standard ISO 8612.2
for tonometers. The study was performed in
the Department of Ophthalmology, Helsinki
University Central Hospital. In the study, 158
patients were measured. The mean paired
dierence and standard deviation (Goldmann-
Icare) were -0.4 mmHg and 3.4 mmHg. A
scattergram and Bland-Altman plot of the
results is shown in Figure 20.
MAINTENANCE
Follow local regulations and recycling
instructions regarding the disposal or recycling
of the Icare tonometer and accessories.
WARNING
The tonometer should only be opened by
qualied service personnel. It contains no
userserviceable parts, apart from the batteries
and a probe base. The Icare tonometer requires
no routine servicing or calibration other than
changing the batteries at least every 12 months
or changing or cleaning the probe base. If
servicing is necessary, contact qualied service
personnel or your local Icare representative.
TECHNICAL INFORMATION
Type: TA011
Dimensions: 24 - 29 mm (W) * 35 - 95 mm (H) *
215 mm (L).
Weight: 140 g (without batteries), 230 g (4 x AA
batteries).
Power supply: 4 x AA non-rechargeable
batteries, 1.5V alkaline LR6.
Measurement range: 7-50 mmHg
Accuracy: ±1.2 mmHg (≤20 mmHg) and ±2.2
mmHg (>20 mmHg).
Repeatability (coecient of variation): <8 %.
Accuracy of display: 1 mmHg.
Display unit: Millimeter of mercury (mmHg).
The serial number is on the inside of the battery
compartment cover.
There are no electrical connections from the
tonometer to the patient.
The device has BF-type electric shock
protection.
Operation environment:
Temperature: +10 °C to +35 °C
Relative humidity: 30 % to 90 %
Atmospheric pressure: 800 hPa-1,060 hPa
Storage environment:
Temperature: -10 °C to +55 °C
Relative humidity: 10 % to 95 %
Atmospheric pressure: 700 hPa-1,060 hPa
Transport environment:
Temperature: -40 °C to +70 °C
Relative humidity: 10 % to 95 %
ACCESSORIES
SKU PRODUCT DESCRIPTION WEIGHT DIMENSIONS
104 Box of 100 probes 89 g 53 x 109 x 36 mm
7217 TA011/TV011 collar 4 g 18 x 18 mm
7218 TA011/TV011 collar narrow 1 g 17 x 18 mm
540 Probe base 4 g 7 x 38 mm
559 Wrist strap with lock 4 g 10 x 10 x 270 mm
525 Aluminium case, Icare ic100 800 g 240 x 280 x 72 mm
7169 Battery cover & screw, Icare ic100 6 g 110 x 25 x 12 mm
623 IOP pad, Icare ic100 38 g 50 x 53 x 16 mm
543 Probe base cleaning container 3 g 20 x 56 mm
565 Silicone grip - white, Icare ic100 26 g 45 x 35 x 113 mm
548 Screw driver, Icare 15 g 16 x 90 mm
577 USB memory stick 44 g 98 x 11 x 93 mm
544 Probe base cover 1 g 19 x 11 mm