GUIDANCE AND MANUFACTURER’S DECLARATION–ELECTROMAGNETIC EMISSIONS
Icare ic100 (TA011) is intended for use in the electromagnetic environment specied below.
The user of the Icare ic100 (TA011) should assure that it is used in such an environment.
RF emissions CISPR 11 Group 1 Icare ic100 (TA011) is battery operated and
use RF energy only for its internal function.
Therefore, its RF emissions are low and are
not likely to cause any interference in nearby
equipment.
RF emissions CISPR 11 Class B Icare ic100 (TA011) is suitable for use in
all establishments, including domestic
establishments and those directly connected to
public low-voltage power supply network that
supplies buildings used for domestic purposes
Harmonic emissions
IEC 61000-3-2
NOT APPLICABLE
Voltage uctuations
ickering emissions
IEC 61000-3-3
NOT APPLICABLE
GUIDANCE AND MANUFACTURER’S DECLARATION–ELECTROMAGNETIC IMMUNITY
Icare ic100 (TA011) is intended for use in the electromagnetic enviroment specied below.
The customers or users of Icare ic100 (TA011)) should assurre that it is used in such an enviroment
Immunity test IEC 60601 Test level Compliance level Electromagnetic environment-
Guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 6 kV contact
±8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or
ceramic tile. If oors are covered
with synthetic material, the relative
humidity should be at least 30 %
Electrical fast
Transients/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/
output lines
NOT
APPLICABLE
NOT APPLICABLE
Surge
IEC 61000-4-5
±1 kV for line(s) to
line(s)
±2 kV for line(s) to
earth
NOT
APPLICABLE
NOT APPLICABLE
Voltage dips, short
interruption and
voltage variations on
power supply lines
IEC 61000-4-11
<5 % UT (>95 %
dip in UT) for
0,5cycle
40 % UT (60 % dip
in UT) for 5 cycles
70 % UT (30 %
dip in UT) for
25cycles
<5 % UT (>95 %
dip in UT) for 5 s
NOT
APPLICABLE
NOT APPLICABLE
Power frequency
(50/60 Hz) magnetic
eld
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic elds
should be at levels characteristic
of a typical location in a typical
commercial or hospital environment.