VF1
Iconic Vacuum Forming Machine
Instruction Manual

Questions and Answers

  • M
    Mary PerezAug 15, 2025
    Why is the motor cabinet (01) of my Iconic VF1 Laboratory Equipment too hot?
    • H
      hdavisAug 15, 2025
      If the motor cabinet (01) of your Iconic Laboratory Equipment is too hot, it might be because the machine has been operating for too long. Try waiting 30 minutes for it to cool down before resuming work. Another possible cause is blocked motor ventilation. Check the space between the support base and the cabinet bottom for any obstructions.
  • H
    Heather CruzAug 18, 2025
    What to do if the vertical movement on the column of the plate fixation unit is very hard on my Iconic VF1?
    • D
      Danielle RobertsAug 18, 2025
      If the vertical movement on the column of the plate fixation unit of your Iconic Laboratory Equipment is very hard, first check if the right cable of vacuum actuation (03) is tight. If so, loosen the handle half way in a counter-clockwise direction. Also, residues or dirt in the upright can cause this issue, so turn off the machine and clean the column with a dry cloth by doing vertical movements with the blank fixation unit. Finally, the machine may be overheated, causing excessive expansion of the sliding elements, wait 30 minutes until the machine cools down and resume work.
  • F
    Frank ColemanAug 21, 2025
    What to do if the blank gets loose from the fixing unit at the plasticization on my Iconic VF1?
    • J
      Jesus RileyAug 21, 2025
      If the blank gets loose from the fixing unit at the plasticization on your Iconic Laboratory Equipment, it is likely due to insufficient tightening of the blank by the lock ring (11). Tighten it a little more and check the blank thickness. The machine accepts blanks of up to 6mm thick.
  • V
    Virginia SerranoAug 25, 2025
    How to fix weak suction on Iconic Laboratory Equipment?
    • R
      Robert SmithAug 26, 2025
      If you are experiencing weak suction, insufficient vacuum forming, or undefined contours with your Iconic Laboratory Equipment, here are a few things to check: * Ensure that the vacuum chamber holes are free by cleaning them with a 1mm Ø metallic pin. * Verify the type of plaster you are using, and avoid resined plaster. * If the model is too high, use the model holder on the cup side or cut the model away. * Check if the cup type model holder (7) is damaged, and replace it if necessary.
  • P
    porteramandaAug 28, 2025
    What to do if the general switch starts but heating does not work on my Iconic VF1?
    • K
      Kayla Williams MDAug 28, 2025
      If the general switch of your Iconic Laboratory Equipment starts but the heating does not work, it is likely that the heating element is burnt. Replace the heating element, call the Authorized Service.
  • G
    Gloria VargasAug 31, 2025
    What to do if general switch starts, but heating and motor don’t on my Iconic VF1?
    • J
      Joel WilliamsAug 31, 2025
      If the general switch of your Iconic Laboratory Equipment starts, but neither the heating nor the motor are working, there may be problems with the internal wiring. Call the Authorized Service to resolve this issue.
  • J
    Julie JonesSep 2, 2025
    What to do if plate fixation unit will not lock out under heating unit of Iconic VF1?
    • A
      Alicia MedinaSep 2, 2025
      If the plate fixation unit of your Iconic Laboratory Equipment will not lock out under the heating unit, the plastic resistance bung may be missing. Undo the right side handle (white dot) and replace the missing plastic resistance bung.
  • L
    laurabakerSep 6, 2025
    What to do if heating is normal but motor does not start on my Iconic VF1 Laboratory Equipment?
    • J
      John HernandezSep 6, 2025
      If the heating is normal but the motor does not start on your Iconic Laboratory Equipment, the end-of-course pin (32) may have some imperfection. Replace or adjust the end-of•course pin (32).
  • A
    andreawalshSep 8, 2025
    What to do if rotary movement of the plate fixation unit does not stop at end of course on my Iconic VF1 Laboratory Equipment?
    • M
      Melissa SanchezSep 9, 2025
      If the rotary movement of the plate fixation unit of your Iconic Laboratory Equipment does not stop at end of course, the rotary pin (29) may be misadjusted. Adjust the pin (29) by the screw (27), guide by the exploded drawing.
  • T
    Thomas MendozaSep 12, 2025
    What to do if rotary movement of the plate fixation unit is locked on my Iconic VF1 Laboratory Equipment?
    • M
      Matthew LindseySep 12, 2025
      If the rotary movement of the plate fixation unit is locked on your Iconic Laboratory Equipment, the rotary lock screw may be too tight. Loosen the screw with the Allen wrench supplied with the equipment.

Summary

INTRODUCTION

TECHNICAL SPECIFICATIONS

PACKAGING CONTENTS

INSTALLATION INSTRUCTIONS

Installation Procedure and Environment

Covers mounting handles, machine placement, electrical safety, and installation area requirements.

Operational Conditions and Requirements

Specifies ideal operating parameters such as temperature, humidity, altitude, and voltage variations.

INSTRUCTIONS FOR USE

Conventional Plastic Forming Process

Step-by-step instructions for performing vacuum forming using the conventional method.

Rotary Plastic Forming Process

Detailed instructions for utilizing the rotary system for vacuum forming on both sides of the blank.

SAFETY AND FUNCTIONALITY PRECAUTIONS

Material and Model Handling for Safe Operation

Advice on plaster selection, model trimming, and heating strategies for various blank types.

MAINTENANCE AND CLEANING

Fuse Replacement Procedure

Step-by-step instructions for safely replacing a blown fuse in the VF1 machine.

SPECIAL STORING CONDITIONS

TROUBLESHOOTING

WARRANTY TERMS AND SERVICING

AUTHORISED SERVICE

Iconic VF1 Specifications