General specifications
Characteristics Description
Display 3½ Digit LCD display
Display Count 2000 count, maximum reading 1999
Overrange Indication “1___” most significant digit
Sampling Rate 3 time/second
Operating Environment:
Relative Humidity
0°C to 40°C (32°F to 104°F)
80% for temperatures up to 31°C
Decreasing linearly to 50% at 40°C
Storage Environment:
-20°C to 60°C (-4°F to 140°F) at <80% relative
humidity
Power source: 9V Battery (NEDA 1604)
Battery Live: 150 hours typical (alkaline)
Low Battery Indicator:
symbol indicates low battery voltage
mA protection Fuse 500mA, 250V fast acting fuse
Dimensions 5.0” H X 2.8” W X 1.4” D
128mmH X 72mmW X 36mmD
Weight: Approximately 7.0 oz. or 200g including battery
RANGES and ACCURACY SPECIFICATION
Function Setting Ranges Accuracy
AC Voltage 200.0/500 V
1.0% ± 4 digits
DC Voltage 2.000/20.00/200.0/1000 V 0.5% ± 2 digits
AC Current n/a n/a
DC Current 2.000mA/20.00mA/200.0mA/
10.00A
1.0% ± 1 digit
2.0%
± 2 digits
Resistance
200.0
2.000K/20.00K/200.0K
2.000M
1.2% ± 3 digits
1.0% ± 1 digit
1.5% ± 2 digits
Diode Check DCV 2V @ 1.0 mA ± 0.6 mA
AC Converter: Average responding, RMS Calibrated to Sine Wave
Overload Protection: AC and DC Voltage: 1100 VDC or 500VAC RMS for no more than one minute.
Resistance: 250VDC or AC RMS
10A input, un-fused
mA input, 500mA at 250V fast acting fuse.
Form number TM61603 Rev 3 February 2002