EasyManua.ls Logo

IDEAL 61-766 - Specifications

IDEAL 61-766
13 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 8
Calibration Procedure (cont’d)
A AC Calibration:
61-764 (Refer to Figure 1B), 61-766 (Refer to Figure 2B) (Adjustments made under front panel label.)
1. Set the function / range to the 660A AC.
2. Set output of the AC calibrator for 10.00A/60Hz +/- 0.01% and connect it to Coil = 10N = 100.0A
AC.
3. Clamp the jaws to the coil = 10N.
4. Adjust VR2 (VR 1KΩ) for a display reading of 100.0 ±1 digit. (61-764)
Adjust VR4 (VR 2KΩ) for a display reading of 100.0 ±1 digit. (61-766)
5. De-energize source and remove test leads.
A DC Zero Calibration
61-768 (Refer to Figure 3B)
1. Set the function/range to A DC.
2. Short the V and COM inputs on the meter.
3. Adjust VR5 (VR 5KΩ) until display reads 000.
4. De-energize source and remove test leads.
A DC Calibration
61-768 (Refer to Figure 3B)
1. Set the function / range to the A DC.
2. Set output of the DC calibrator for +10.00A ± 0.01% and connect it to Coil = 10N = +100.0A DC.
3. Clamp the jaws to the coil = 10N.
4. Adjust VR4 (VR 3.3KΩ) for a display reading of 100.0 ±1 digit.
5. De-energize source and remove test leads.
Calibration of the 61-760 series is complete.
Remove all leads from the calibrator and equipment.
Return unit to proper operating condition.
Form number TM61764-6-8 Rev 5 November 2007

Other manuals for IDEAL 61-766

Related product manuals