EasyManua.ls Logo

IFM Electronic SU8000 - Advanced User Settings

IFM Electronic SU8000
31 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
29
UK
10.5.7 Set measured value damping
Select [dAP] and set the damping constant in seconds (τ value 63 %).
10.5.8 Set output status in fault condition
Select [FOU1] and set the value:
- [On] = output 1 switches ON in case of an error
- [OFF] = output 1 switches OFF in case of a fault�
Select [FOU2] and set the value:
- [On] = output 2 switches ON in case of a fault, the analogue signal
goes to the upper final value�
- [OFF] = output 2 switches OFF in case of a fault, the analogue signal
goes to the lower final value�
10.5.9 Select the medium to be monitored
Select [MEDI] and set the requested medium:
- [H2O] = water
- [GLYC] = glycol solutions
- [OIL�1] = High viscosity oil (viscosity: 30���68 mm²/s at 40°C)
- [OIL�2] = Low viscosity oil (viscosity: 7���40 mm²/s at 40°C)
10.6 Service functions
10.6.1 Read min/max values for volumetric flow
Select [HI] or [Lo], briefly press [Set]�
[HI] = maximum value, [Lo] = minimum value�
Delete memory:
Select [HI] or [LO]�
Press and hold [Set] until [----] is displayed�
Briefly press [Mode/Enter]�
It makes sense to delete the memories as soon as the unit operates under
normal operating conditions for the first time�
10.6.2 Reset all parameters to factory setting
Select [rES]�
Press and hold [Set] until [----] is displayed�
Briefly press [Mode/Enter]�
For the factory settings please refer to the end of these instructions → 13
We recommend taking down your own settings in that table before carrying
out a reset�

Related product manuals