EasyManua.ls Logo

IFM PI2898 - Customer-Specific Calibration

IFM PI2898
33 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8
Factory setting Measuring range scaled









P = system pressure , MAW = initial value of the measuring range, MEW = final value of
the measuring range
1
: [OU2] = [I];
2
: [OU2] = [InEG]
In the set measuring range the output signal is between 4 and 20 mA ([OU2] = [I])
or between 20 and 4 mA ([OU2] = [InEG]).
It is also indicated:
System pressure above the measuring range:
- Output signal > 20 mA at [OU2] = [I].
- Output signal 4 to 3.8 mA at [OU2] = [InEG].
System pressure below the measuring range:
- Output signal 4 to 3.8 mA at [OU2] = [I].
- Output signal > 20 mA at [OU2] = [InEG].
4.4 Customer-specific calibration
The customer-specific calibration changes the curve of measured values
compared to the real measured values (shifting / change of the gradient;
→ 9.4.6 [CAL]).
Two calibration points can be defined (CP1, CP2). The two points are
independent of each other.
The two calibration points must be within the scaled measuring range
(→ 4.3 ).
The zero point calibration [COF] influences the calibration of the curve of
measured values. Recommendation: Set [COF] to 0 (→ 9.4.1 [COF]), then
calibrate the measured value curve.
After a change the calibration can be reset to factory setting (→ 9.5.2 [rES]).

Related product manuals