IGEA OSTEOSPINE – User Manual
IGEA OSTEOSPINE – User Manual EN
23
the result of an improper use. However there is no other kind of risk as the result of over-charging
except damage to the same.
3.7 Generator cleaning
The device must be used in compliance with the normal rules of hygiene, and should be cleaned
regularly. The presence in the environment of use of hair, hair, dust and exposure to direct sunlight,
even if it does not cause malfunction of the device, should be avoided.
Before cleaning the generator ensure that the device is switched off and disconnected from the
battery charger: the generator can be cleaned with a damp cloth and neutral detergents, never use
any aggressive detergent or solvent agents to clean the generator. Electrodes do not need cleaning
and can be used for more treatment sessions.
4. TROUBLE SHOOTING
4.1 Error messages
The device notifies the fault conditions; the table below shows the reports provided and the actions
required to restore correct operation.
On positioning the switch to “on” (I) the yellow “ ”
indicator fails to come on.
The internal power battery may be
completely flat, recharge it as
described in paragraph 3.6
The stimulator is on with only the yellow “ ”indicator light
on in a continuous manner, the buzzer emits an intermittent
sound which is continually repeated.
The electrodes have not been applied
to the patient or they have become
disconnected from the generator.
Check the correct positioning and
connection. If the message persists,
the cable might be broken and must be
Even if the electrodes are correctly connected, the yellow “
”indicator light is on in a continuous manner.
Check that electrodes or cables are not
in direct contact with each other.
The stimulator is on and in operation, the buzzer emits a
sound and the red “ ”indicator begins flashing.
The stimulator is operating correctly,
but the battery is low, from the time of
the signal the appliance has a residual
autonomy of some hours before
running out. The battery must be
recharged as soon as possible.
The stimulator is on with yellow “ ” indicator flashing,
continuous red “ ” indicator light on and buzzer emitting
a “beep” every 5 seconds for a total of 30 seconds; before
the device goes off.
The stimulator is on and operational,
but the battery is totally flat and can no
longer power the circuit. The battery
needs to be charged before resuming
treatment.
This condition is preceded by the
condition described above.
The stimulator is on with continuous yellow “ ” and red “
” lights on and buzzer emitting a “beep” every 5
seconds for a total of 30 seconds; before the device goes off.
The generator founds irregular; the
device does not provide any signal and
maintenance service is required.