Input window NEW MEASUREMENT
Select and confirm the menu item “Sample weight“. Enter the
sample weight with numeric keypad and confirm.
INFO:
A distinction is made between the use and non disposable
cup. Use of disposable crucible can be selected in Settings
" Accessories.
INFO: For automatic input of the weight may be a connected
scale can be used
INFO: The inputted weight values are checked for permissibility.
INFO: The existing store will be reviewed after the entry of a
new measurement / calibration and a user note displayed
if it is not possible to save the measurement / calibration.
NOTE:
Ensure that a maximum energy input of from 40,000 joules is not
exceeded.
NOTE:
For more, detailed descriptions of the individual menu items are
treated in the chapter “Calorimetric Measurements“.
INFO: If a scale is used in activation of the input field, the
weight is automatically queried. It can be pressed on the
scale alternatively, the transmit button.
Fill in the sample weight.
Calibration off.
When you are ready with your input, press “OK” button.
Don’t close the C 1 before you have pressed “OK“
button!
During the prerequisite checking, the cotton thread can be
attached to the ignition wire, hung the crucible in the provided
holder and the sample are contacted with the ignition wire.
Placing the sample
When using a disposable crucible of contact between the ignition
wire and the combustible crucible must be present. The fixing of
the cotton thread is not necessary.