EasyManua.ls Logo

IKA C 1 - Maintenance Procedures Menu; Maintenance Menu Functions

IKA C 1
44 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
32
Selection of accessories
Heating:
Use of heating during the measurement C 1.2
Combustible crucible:
yes / no
Weight input of the combustible crucible in the submenu “New
measurement“ can be done manually or by data transfer from
the balance.
Printer:
Using a serial printer.
Selection option between:
• off: no expression
• briefly: expression
• standard: result expression with temperature values
Service Information:
Service information is outputted via the serial interface
Balance:
Selecting a balance type.
Selection option between:
out
Sartorius/Kern
Mettler-Toledo
Ohaus
Accessories
Chiller test
Viewing maintenance programs that can be performed by the
user at regular intervals.
Inner vessel and the inner vessel bottom lift out and set
aside.
Maintenance menu:
The system must be open to performing the maintenance programs.
The maintenance menu can be reached at Menu > Maintenance and contains the following points
Check whether the cooling water tem-
perature within the specified range of
19.5 +/- 1.5 °C at the wor-
king temperature 22 °C
or
27.5 +/- 1.5 °C at the wor-
king temperature 30 °C
is located.
Maintenance

Table of Contents

Other manuals for IKA C 1

Related product manuals