33
Change seal:
1 In the Menu Maintenance press “change seal“ button. This
requires oxygen pressure be present at C 1.
2 Unscrew the piston (Pos. 38 III) with screwdriver
counterclockwise
3 Remove the O-ring (Pos. 38 VI)
4 Put new O-ring on piston (Pos. 38 VI)
5 Screw in piston and tighten securely
6 “Change seal“ button again actuate pistons move to the
home position
38
38
III
Stirrer test
The stirrer is switched on and off for a visual check.
O
2
System blow:
1 Lift out the inner vessel from the bottom and set aside
2 After about 20 seconds the button “O
2
system blow“ button.
Keep openings with a paper towel or similar covered by hand
3 After about 3 seconds, this state switches itself off again.
This any remaining condensed water and dirt is
in the line blown out.
This screen shows:
the software and firmware version
the serial number of the device
the date of the last service
the date of the last pressure test
the number of performed ignitions
the number of ignitions in the next pressure test must be
performed
the number of performed ignitions and support information
IKA
®
- Contact information
blow
O-ring
(Water inlet valve
seal)
Stirrer
Servo up
The locking pin is extended.
Servo down
The locking pin is retracted.
VI (8,0 x 2,5)