Commissioning
C 6000
C 6000
Condenser
C 6000
C 6000
• O
2
connection tube
(in)
• Inflow pipe
(in)
• Venting hose
(out)
• Discharge hose
(empty)
M 8x1
Place of installation
A constant ambient temperature is an important
requirement for ensuring the high measuring
accuracy of the system. The following conditions
must therefore be fulfilled at the place of
installation:
•
No direct solar radiation
• No draughts (e.g. beside windows, doors, air
conditioning)
• Sucient distance to radiators and other heat
sources
• The minimum distance between the wall and
the rear side of the device must be at least 25
cm.
• The system must not have laboratory hardware
such as shelves, wire ducts, ring lines, etc.
installed above it.
• The (constant) room temperature should be
around 20 °C ... 25 °C.
• The system must be set up on a horizontal
surface.
For operation of the system the following must be
available at the place of installation:
- A power supply corresponding to the nameplates
on the system components,
- An oxygen supply (99.95 % pure oxygen, quality
3.5; pressure 30 bar) with pressure gauge.
A shut-off valve for the oxygen supply must be
installed. Observe the instructions on handling
oxygen given in the chapter „Safety instructions“.
Please read these operating instructions carefully.
IKA
®
consider themselves responsible for the
safety, reliability and performance of the appliance
only when
• the device has been operated in accordance
with the operating instructions,
• the specifications for the place of
installation have been met.
Do not use distilled or demineralized water
(increased risk of corrosion)! Fill the cooler
according to the cooler operating instructions. is
recommended in drinking tap water quality. Mix
in (max. 1 ml for 4-5 l of water) of the supplied
water additive. Thus, the durability of the water is
improved.
• Return pipe
(out)
Hoses