Display Description
What should be done?
What happens next?
Water not filling OK. The system could not completely
fill the inner vessel with
water completely within the spe-
cified period of time.
Acknowledge the error message
to revert to the „Wait“ message.
Water level in the condenser too
low.
Water flow rate insucient.
Defect at the water sensor. The water sensor shows comple-
tely full within a very short time.
Acknowledge the error message
to revert to the „Wait“ message.
Check whether a drip is handing
from the water sensor.
Blow out the recess around the
water sensor.
Rfid ID of the new decompositi-
on vessel not accepted.
This error message is triggered
by the dialogue „New decom-
position vessel“ if you attempt
to register a new decomposition
vessel at the C6000 using an Rfid
ID that has already been used.
Acknowledge the error message
to continue the dialogue.
Use a different Rfid ID for this de-
composition vessel.
Contact IKA
®
Customer Service.
xxxxxxxxxxx : This Rfid ID cannot
be assigned
The Rfid value detected does not
match any of the decomposition
vessel registered at the C 6000.
Acknowledge the error message
to revert to the „Prepared“ mes-
sage.
Check whether the vessels have
been mixed up.
Check whether a misleading Rfid
is stated anywhere.
Register this decomposition ves-
sel at the C 6000.
Converter alarm High-precision temperature
measurement is one of the core
functionalities of the calorimeter.
Any error message regarding the
temperature recording brings
the current measurement to an
end and demands a system test.
Further work using the calorime-
ter can be performed only once
the system test has been com-
pleted successfully.
Acknowledge the error message
in order to switch to the „System
Check“ status.
Press the „Restart“ button“.
The temperature converter is re-
set and the system test starts.
On successful completion the
status switches to „Wait“
If the system test was unsuccess-
ful, switch the C 6000 off and on
again.
If the system test is unsuccessful
again, please contact IKA
®
Cus-
tomer Service.