55
Various major parts of the calorimeter are mo-
nitored. Information about when the servicing
intervals defined by the manufacturer for these
parts fall due is shown in the information field,
with the invitation to perform servicing actions.
You must then confirm performance of the ser-
vicing actions.
Click on „Confirm“ to acknowledge performance
of the servicing actions.. The acknowledgement
is saved and can be checked by the Service En-
gineers.
If you click on the „Help“ button you will be given
instructions on exactly what to do.
This monitoring is the precondition for a high
degree of safety and we ask you to follow the
stated instructions precisely.
At the latest after 500 measurements, check the
O-ring, filter etc..