Technical data
IKA
®
C 6000 global standards
Measuring range max. ..................................................................................... J ................................................ 40000
Measuring mode adiabatic 22 °C .............................................................................................................................yes
Measuring mode dynamic 22 °C ...............................................................................................................................yes
Measuring mode isoperibolic 22 °C ..........................................................................................................................yes
Measuring mode adiabatic 25 °C .............................................................................................................................yes
Measuring mode dynamic 25 °C ..............................................................................................................................yes
Measuring mode isoperibolic 25 °C ..........................................................................................................................yes
Measuring mode adiabatic 30 °C .............................................................................................................................yes
Measuring mode dynamic 30 °C ..............................................................................................................................yes
Measuring mode isoperibolic 30 °C ..........................................................................................................................yes
Measuring/hour adiabatic ........................................................................................................................................... 5
Measuring/hour dynamic ............................................................................................................................................ 7
Measuring/hour isoperibolic ........................................................................................................................................ 4
Reproducibility adiabatic (1g benzoic acid NBS39i) ....................................... %RSD ............................................... 0.05
Reproducibility dynamic (1g benzoic acid NBS39i) ........................................ %RSD ............................................... 0.15
Reproducibility isoperibolic (1g benzoic acid NBS39i) .................................... %RSD ............................................... 0.05
Touchscreen .............................................................................................................................................................yes
Working temperature min. .............................................................................. °C .................................................... 22
Working temperature max. .............................................................................°C .................................................... 30
Temperature measurement resolution ..............................................................K .............................................. 0.0001
Cooling medium temperature min. ..................................................................°C .................................................... 12
Cooling medium temperature max. .................................................................°C .................................................... 27
Cooling medium permissible operating pressure ............................................. bar ................................................... 1.5
Cooling medium .......................................................................................................................water quality tap water
Type of cooling ....................................................................................................................................continuous flow
Flow rate min. .................................................................................................l/h .................................................... 60
Flow rate max. ................................................................................................ l/h .................................................... 70
Recommended flow rate at 18 °C ...................................................................l/h .................................................... 60
Oxygen operating pressure max ..................................................................... bar .................................................... 40
Interface to scales ......................................................................................................................................RS232 serial
Interface to printer ................................................................................................................. USB, RS232 and network
Interface to PC ....................................................................................................................................................RS232
Interface to sample rack .................................................................................... .....................................................yes
Interface to ext. keyboard ........................................................................................................................................yes
Oxygen filling .................................................................................................... .....................................................yes
Degassing ................................................................................................................................................................yes
Decomposition vessel detection ...............................................................................................................................yes
Decomposition vessel C 6010 ..................................................................................................................................yes
Decomposition vessel C 6012 ..................................................................................................................................yes
Evaluation to DIN 51900 ..........................................................................................................................................yes
Evaluation to DIN EN ISO 1716 ................................................................................................................................yes
Evaluation to DIN EN ISO 9831 ................................................................................................................................yes
Evaluation to DIN EN 15170 .....................................................................................................................................yes
Evaluation to DIN CEN TS 14918 ..............................................................................................................................yes
Evaluation to ASTM D240 ........................................................................................................................................yes
Evaluation to ASTM D4809 ............................................................................... .....................................................yes