/// Menu details
Operating mode:
Mode A:
All settings will be stored if the device is switched off or disconnected from the power supply
The stirring and heating functions will be set to OFF when the device is powered on� The safety
circuit can be set or modified� When the power supply switch is switched on, A is displayed�
Mode B:
All settings will be stored if the device is switched off or disconnected from the power supply
The stirring and heating functions will be set to ON or OFF when the device is powered on,
depending on the previous status of the device� The safety circuit can be set or modified� When
the power supply switch is switched on, B is displayed�
Mode D:
When switching the device on, the safe temperature has to be confirmed by pushing one of
the rotating / pressing knobs�
During operation, in mode D the device behaves the same as in Mode A with the exception
that:
The temperature / speed setting must be confirmed by pressing the rotating / pressing knob�
To change the temperature / speed setting, turn the rotating / pressing knob until the desired
value appears�
The new value flashes for 5 seconds on the display� Confirm the new temperature / speed
setting by pressing the rotating / pressing knob, otherwise the temperature / speed setting
value will skip back to its previous value�
Temperature control mode:
When using an external PT 1000 temperature sensor, you can choose below control modes:
PID mode:
Good control results, minimized overshooting, slow rise in temperature�
2P-mode (two-point controller):
Maximum heating rate, increased overshooting�
PI mode:
Good control results, for some of the heating blocks�
T
Set
t
PID
T
Set
t
2P
Temp. control
Factory settings
Intermittent mode /
Rotation direction
Intermittent mode
Rotation direction
Automatic reverse rotation
Run time
Pause time
00:00 [mm:ss]
00:03 [mm:ss]
Clockwise (continuous)
ON
OFF
-
Beep
ON
OFF
-
Temperature unit
°C
°F (not available for Japan)
-
Timer expired
yes
no
yes
no
Stop heating
Stop stirring
-
-
Safe / Visc
Safe
Visc
-
SAFE
Factory reset
yes
no
-
Probe calibration
(not default) reset
no
yes
(default) set
no
yes
2 Point
-
-
Software version
-
Error 5
05:00 [mm:ss]
Operating mode
A
B
D
-
-
Temperature control mode
PID
2P
PI
-
-
/// Menu structure