EasyManua.ls Logo

IKA Plate - Technical Data

IKA Plate
21 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
36 37
Technical data
IKA Plate (RCT digital)
RCT 5 digital
RCT digital
(650 W)
RCT digital
(900 W)
RCT digital
White
General data
Voltage 220 – 230 VAC ± 10 %
115 VAC ± 10 %
100 VAC ± 10 %
Frequency 50 / 60 Hz
Power input 650 W 900 W
Power input standby 1�6 W
Self-heating of the heating plate by max�
stirring (RT:22° C / duration:1h)
+ 13 °C
Automatic reverse rotation yes
Intermittent mode yes
Viscosity trend measurement yes
Timer yes
Interface USB, RS 232
Permissible ambient temperature + 5 ��� + 40 °C
Permissible relative humidity 80 %
Protection class according to DIN EN 60529 IP 42
Protection class I
Contamination level 2
Overvoltage category II
Plate material Aluminium alloy Aluminium with ceramic
coating
Plate dimensions Ø 135 mm
137 x 137 mm
Dimensions (W × D × H) 160 × 270 × 85 mm
Weight 2�3 kg 2�5 kg 2�5 kg 2�6 kg
Operation at a terrestrial altitude max� 2000 m
Stirring function
Number of stirring positions 1
Max� stirring quantity (H
2
O) 20 l
Speed range 0 / 50 ��� 1500 rpm
Speed display set-value LCD
Speed display actual-value LCD
Speed setting rotating / pressing knob
Speed setting accuracy 10 rpm
IKA Plate (RCT digital)
RCT 5 digital
RCT digital
(650 W)
RCT digital
(900 W)
RCT digital
White
Speed variation (no load, nominal voltage,
at 1500 rpm, ambient temperature + 25 °C)
± 2 %
Stirring bar length 30 ��� 80 mm
Heating function
Heat output 600 W 850 W
Heating temperature range RT + device self-heating ��� 310 °C
Temperature setting range 0 ��� 310 °C
Temperature display set-value LCD
Temperature display actual-value LCD
Temperature setting rotating / pressing knob
Temperature setting resolution of heating
plate
1 K
Temperature setting resolution of medium 1 K
Heating rate (1 l water in H 1500) 6�5 K / min 8�5 K / min
Temperature control accuracy of heating
plate (at 100 °C)
± 5 K
Adjustable safety circuit 50 °C ��� 370 °C
External temperature sensor / thermometer
Connection for external temperature
sensor / thermometer
PT 1000 series (excluding double PT 1000 sensors),
ETS-D5, ETS-D6
Temperature control accuracy (500 ml
water in 600 ml glass beaker, 40 mm bar,
600 rpm, 50 °C)
± 0�5 K (with temperature sensor PT 1000)
± 0�5 K (with thermometer ETS-D5)
± 0�2 K (with thermometer ETS-D6)
Temperature sensor PT 1000 deviation
EN 60751 class A
≤ ± (0�15 + 0�002 × ITI)
Sensor in medium detection (Error 5) yes