22
Error codes
If an error occurs, this is shown in an error on the display (B), e.g. Error 4.
Proceed as follows:
• Switch the device off on the device switch,
• Take corrective measure,
• Switch device back on.
Error code Description Detection Reason Action
Error 3 Device temperature
error
The temperature of
device has exceeded the
limit.
The temperature is too high. Stop the device for a while and restart
again. If the problem occurs again and
again, please call service department.
Error 4 Motor overload Motor overload The motor is blocked be-
cause of overload.
Stop the device for a while and restart
again. If the problem occurs again and
again, please call service department.
Error 8 Speed sensor fault. Device can’t detect the
pump speed.
The sensor occurs some
unknown errors that device
can’t read the speed value.
Call service department.
Error 9 Storage Error Read or write internal
flash error.
Read or write internal flash
error.
Call service department.
In the event that the fault cannot be eliminated using the measures described or another error code is displayed:
• Please contact the Service Department,
• Send the device and a brief description of the fault.
Accessories
• Vacuum hose IKA VH.SI.8
• PC 1.1 cable (RS 232)
• Woulff bottle/water valve set VSW1
For further accessories see www.ika.de
Product contact parts
Name Material
Connection spigot PPS
Connection piece PPS
Distributor PPS
Head piece PPS
Diaphragm valve/O-ring FFPM
Diaphragm NBR/PTFE
Clamp PPS
Connection piece PTFE
Pressure sensor AL203