iM3 Dental Systems
Operation Recommendations
Air pressure The dental unit air pressure
should be adjusted to be 40
psi See page 16
Lubrication To assure optimal
performance and life the
handpiece should be
lubricated daily with MD-30 oil
lubricant. Instructions
for lubrication are found on
page 20.
Air supply Moist or dirty air from the
compressor can cause
damage. Check the regulator
filter regularly. If dirt or
moisture is present, replace
the regulator filter and bleed
the system, to remove water
or oil.
Installation Instructions
1. Preset the air pressure to 40 psi
(see page 16)
1. Connect the low speed handpiece to
the grey handpiece line
2. Fit the 4:1 Advantage straight nose
cone. Attach the Blue R&R oscillat-
ing prophy head
3. Depress the foot control fully to oper-
ate the low speed handpiece
Note: Polishing should be carried out at 5,000 rpm or less. This can be
accomplished by running the motor at full revs (20,000 Rpm) and use a 4:1
reduction head, to achieve 5,000 revs, maintaining torque.
iM3 Advantage Low Speed Handpiece L6200
The low speed handpiece features include a light weight balanced profile and
smooth swivel action. Its vane type power system is quiet and vibration free. The
motor accommodates both doriot or ISO heads and its powerful torque is suitable
for all low speed procedures. The Low speed Motor operates at 20,000 RPM at a
delivery pressure of 40 psi.