Handpiece Trouble Shooting
Low Speed Handpiece
PROBLEM CAUSE SOLUTION
Handpiece
won’t run
1. No or low air pressure.
2. Air line improperly sealed.
3. Inferior oil; too much or excess
more commonly too little oil or
water in the turbine.
4. Broken drive head.
Note new handpieces require a
“Run in time” this can be up to 20
hours of use
1. Check the air line to be sure
the air is turned on and the
line is unobstructed.
2. Reconnect air line to
handpiece making sure the
gasket is fully contracted.
3. See lubrication instructions
on page 17.
4. Contact your local iM3
official Service Centre.
If the motor stops spinning
place a fine screw driver down
the nose and rotate shaft.
Low Speed Handpiece - 4:1 Straight Nose Cone
PROBLEM CAUSE SOLUTION
Cant fit R&R pro-
phy angle
1. Plastic drive shaft broken off
internally in Chuck housing.
2. Improper lubrication.
Only fit plastic iM3 R&R prophy
angles Code L7465.
1. Contact your local official
iM3 Service Centre.
2. See lubrication instruc-
tions, on page 17.