Elite LED Dental System
Operating Manual
iM3 Inc iM3 Pty Ltd
12414 NE 95th Street 21 Chaplin Drive
Vancouver WA 98682 USA Lane Cove NSW 2066 Australia
Tel (360) 254 2981 Fax (360) 254 2940 Tel (02) 9420 5766 Fax (02) 9420 5677
www.im3vet.com
Email: enquiries@im3vet.com
2013

Questions and Answers

  • A
    Alejandro MasonAug 12, 2025
    What to do if there is no water in my iM3 Elite handpiece?
    • E
      espinozadorothyAug 12, 2025
      If there's no water flow in your iM3 Medical Equipment handpiece, start by checking a few basic things. Is the compressor turned on? Is the water flow control (P4) in the ON position? Is the water/CLS bottle empty? Look for any kinks in the water line that might be blocking the flow. Make sure the 3-way syringe has water. Check the lines and pressure to ensure they're adequate. Also, verify that the regulator pressure is set to 70-80 PSI.
  • K
    Kimberly AllenAug 22, 2025
    What causes no water flow in iM3 Elite handpiece?
    • K
      Karen RossAug 22, 2025
      If you're not getting water flow in your iM3 Medical Equipment handpiece, it could be due to mineral accumulation from using tap water, which typically occurs over 3-4 years. Another cause could be the use of Chlorhexidine, which can deposit and block valves. CLS is designed for use in its diluted strength without depositing, so ensure you are using it correctly. Other possible causes include the compressor not being ON, the water flow control not being in the ON position, or an empty water/CLS bottle. Also, check for kinks in the water line, ensure the 3-way syringe has water, and verify the lines, pressure, and regulator pressure (70-80 PSI).
  • S
    Sarah RossNov 1, 2025
    Why is there no water flow in my iM3 Medical Equipment handpiece?
    • D
      Denise HarrisNov 1, 2025
      If you're experiencing no water flow in your iM3 Medical Equipment handpiece, several factors could be the cause. Start by ensuring the compressor is turned ON and the water flow control is in the ON position. Verify that the water/CLS bottle is filled. Inspect the water line for any kinks that might be obstructing the flow. Also, check if the 3-way syringe has water and that the lines and pressure are adequate. Confirm the regulator pressure is set between 70-80 PSI. Mineral accumulation from using tap water or deposits from Chlorhexidine can also cause blockages. If it is mineral accumulation, note that this will usually take 3-4 years to occur. Most brands of Chlorhex will deposit out of solution and block valves; CLS is designed to be used in its diluted strength without depositing. To...

Summary

Unique Design Features

Safety and Hygiene Features

Covers CLS Enviro System, Filters, Suction, and LED Handpiece hygiene.

Getting Started

iM3 Distilled Water and CLS Enviro Solution Supply

Refilling Coolant Bottles

Instructions for refilling distilled water and CLS solution bottles safely.

High Speed Handpiece - LED Advantage HS HP L5200 s

Bur Replacement - iPush Tool

Instructions for fitting and using the iPush tool to replace burs.

iM3 Handpiece Pressure Adjustment

Handpiece Pressure Adjustment Procedure

How to adjust pressure for low and high speed handpieces using knurled knobs.

Lubrication of HS & LS Handpieces

Handpiece Lubrication Procedure

Recommends MD-30 oil, application via air entry hole, and running pedal.

Silent Hurricane Compressor - Oil Free

Operating the Silent Hurricane Oil Free Compressor

Steps to connect, switch on, and operate the compressor.

Identifying Air Leaks

Procedure to check for air leaks to prevent motor overheating.

Silent Hurricane Compressor - Adjusting Regulator Air Pressure

Adjusting Regulator Air Pressure

How to adjust the regulator knob to set the desired air pressure.

Handpiece Trouble Shooting

High Speed Handpiece Issues

Common problems like no water flow, mineral accumulation, and chlorhexidine use.

Low Speed Handpiece Troubleshooting

Handpiece Wont Run

Causes and solutions for the low speed handpiece not running.

iM3 Elite Specifications

General IconGeneral
BrandiM3
ModelElite
CategoryMedical Equipment
LanguageEnglish