OPERATOR'S MANUAL
EN
1.2. INTENDED USE
This x-ray unit is designed for use in the dental surgery to make endo-oral x-rays for diagnostic purposes.
This equipment can be used to produce traditional x-rays developed using chemicals or, alternatively, it can be used
with digital x-ray sensors.
WARNING!
This X-ray device is for VETERINARY USE ONLY.
This X-ray device is NOT TO BE USED ON HUMANS.
1.2.1. CLASSIFICATION
- RADIO EQUIPMENT AND TELECOMMUNICATIONS TERMINAL EQUIPMENT Classification.
Equipment classification according to Directive 99/05/EC Art.12: CLASS I.
- EMC classification.
Equipment classification in accordance with standard CEI EN 55011: GROUP I TYPE B.
1.2.2. ENVIRONMENTAL CONDITIONS
The equipment is to be installed in rooms that satisfy the following requirements:
- Temperature from +10 to +40°C.
- Relative humidity from 25 to 75% without condensate.
- Atmospheric pressure from 700 to 1060 hPa.
- The electrical wiring in the room in which the equipment is installed must conform to I.E.C. 60364-7-710;V2
specification (i.e. the regulations concerning the electrical wiring to be used in surgeries) or equivalent standards
in force in the country where the equipment is installed.
- ELECTRICAL CONNECTIONS: The electrical wiring must have an effective ground conductor as set forth by
I.E.C. -US National Electrical Codes and C.E.I. standards. In Italy, electrical wiring must comply to standards IEC
60364-7-710 which require that a ground fault circuit interrupter is installed upstream.The ground fault circuit
interrupter must be as specified below:
- contact capacity: 250V 10A in compliance with standards IEC 60898-1 and IEC 60947-2;
- differential sensitivity: 0.03A;
- power supply: 3x2.5 mm
2
.
The color of the 3 wires should be as specified in the standards (brown power, BLUE neutral, YELLOW/GREEN
ground).

Table of Contents

Other manuals for iM3 REVOLUTION 4DC

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the iM3 REVOLUTION 4DC and is the answer not in the manual?

iM3 REVOLUTION 4DC Specifications

General IconGeneral
BrandiM3
ModelREVOLUTION 4DC
CategoryMedical Equipment
LanguageEnglish

Related product manuals