28
Register
Address:
Decimal
[hex]
Class & Type Function: Binary Format:
45121
[1400]
Class:
Configuration
Type:
Read\write
Display blank 000000000000000b
LSB(b)
0 = disabled
1 = enabled [display will blank]
45122
[1401]
Class:
Configuration
Type:
Read\write
Menu mode 000000000000000m
LSB (m)
0= numeric
1=alphanumeric
45155
[1422]
Class:
Configuration,
AlarmClearing
Type:
Read\write
Auto reset 27 0000000 e
LSB (e
)
0 = disabled
1 = enabled
45156
[1423]
Class:
Configuration,
AlarmClearing
Type:
Read\write
Auto reset 25 0000000 e
LSB (e)
0 = disabled
1 = enabled
45633
[1600]
Class:
Configuration
Type:
Read\write
RS-232 mode 0000000000000rrr
LSBs (rrr)
000 = Serial disabled
001 = datalogger mode
010 = polled mode
011 = sampled mode
100 = command
101 = reserved N/A
110 = RS485 Modbus
111 = remote display driver
45634
[1601]
Class:
Configuration
Type:
Read\write
Baud 000000000000bbbb
LSBs (bbbb)
0000 = 300 0001 = 1200
0010 = 2400 0011 = 4800
0100 = 9600 0101 = 14400
0110 = 19200 0111 = 28800
1000 = 38400 1001 = 57600
1010 = 76800