2.7 Safe Use of the Active Accessories
▪ Examine the connection of the Handpiece through the connector to the System
before using. Improper connection may result in arcs and sparks, accessory
malfunction, or unintended treatment effects.
▪ Do not wrap the Handpiece cords around metal objects. It may induce current that
could lead to electrical shock, fire or injury to the patient or personnel.
▪ Ensure that return electrode is in full contact with the skin. Bad coupling of the
return electrode with the skin results in a specific warning sound, a message on the
screen, and disabling of RF.
▪ Handpieces are for single use only. Do not try to reuse the handpiece. Re-use of the
handpiece may create loss of integrity of handpiece components that may affect the
performance and make the device non-functional.
Do not connect a wet accessory to the System.
Do not immerse the applicator under water at any time.
The Handpiece is gamma-sterilized for a single use only and CANNOT be
autoclaved or re-sterilized by any other technology.
2.8 Warnings
This equipment is for use only by trained, licensed physicians.
Only Handpieces manufactured or approved by InMode MD Ltd. should be
used with the BodyTite System.
Connect the power cord to a properly polarized and grounded power source
with the frequency and voltage characteristics that match those listed on the
back of the unit.
Connect the System power cord to a properly grounded receptacle. Do not use
power plug adapters.
Always turn off and unplug the device before cleaning.
The patient should not come into contact with metal parts which are earthed
or which have an appreciable capacitance to earth. The use of antistatic
sheeting is recommended for this purpose.