9.2 EMC Safety for the BodyTite Device
The device has been tested and found to comply with the limits for the medical devices
to the IEC 60601-1-2. These limits are designed to provide reasonable protection against
harmful interference in a typical clinical installation. This device generates uses and can
radiate radio frequency energy. If not installed and used in accordance with the
instructions, may cause harmful interference to other devices in the vicinity. However,
there is no guarantee that the interference to other devices, which can be determined by
turning the device off and on, is caused by this instrument.
The user is encouraged to try to correct the interference by one or more of the following
measures:
▪ Reorient or relocate the receiving device.
▪ Increase the separation between the devices.
▪ Connect the device into an outlet on a circuit different from that which was
previously used.
▪ Consult InMode MD Ltd. service personnel for help.
▪ Interference to the device may be caused by portable and mobile RF communication
equipment. In case of an interruption, beware of such a device in the vicinity.
▪ Use of the system with any accessory, transducer or cable other than those specified
may result in increased EMISSIONS or decreased IMMUNITY than those specified.
▪ The system should not be used adjacent to or stacked with other equipment. If
adjacent or stacked use is necessary, the system should be observed to verify normal
operation in the configuration in which it will be used.